2026-02-02 14:11:36,753 INFO Start Time: 2026-02-02 14:11:36.753001
2026-02-02 14:11:36,753 INFO [MOCK] Serial terminal initialized
2026-02-02 14:11:36,753 INFO [MOCK] Serial terminal connected
2026-02-02 14:11:36,754 INFO [MOCK] Simulated RECYCLE event: %%F-10747-SP*: ***** RECYCLE!******
2026-02-02 14:11:36,754 INFO The Test Operator check if the failure BIT in B6_MsgRdrSettingsAndParametersTellback changes ...
2026-02-02 14:11:36,754 INFO [MOCK] Log started at level 3 in directory: C:\src\____GitProjects\SXXXXXXX_PyBusMonitor1553\__OLD\__TEST_GENOVA\GrifoAutomaticTestEnv\TestEnvironment\scripts
2026-02-02 14:11:36,756 INFO tgt_gen_alone(): 0
2026-02-02 14:11:36,756 INFO tgt_gen()
2026-02-02 14:11:36,780 WARNING [MOCK] getSingleMessageReceivedSz called but interface not started! Returning 0
2026-02-02 14:11:36,780 INFO TgtMsg: 0 0
2026-02-02 14:11:36,780 INFO Tgt: 1 @ 1180
2026-02-02 14:11:36,780 INFO Check that B9 b9_t_num is between 1 and 2
2026-02-02 14:11:36,780 INFO OK: read value = 1
2026-02-02 14:11:36,780 INFO Check that B9 b9_t1_rng is between 1179 and 1186
2026-02-02 14:11:36,780 DEBUG [MOCK] Field rng: 1180
2026-02-02 14:11:36,782 INFO OK: read value = 1180
2026-02-02 14:11:36,782 INFO [MOCK] Log stopped
2026-02-02 14:11:36,782 INFO Repetition 1 of 3
2026-02-02 14:11:36,782 DEBUG [MOCK] Serial statistics reset
2026-02-02 14:11:36,782 INFO The test operator is required to switch off the target and wait 3 seconds.
2026-02-02 14:11:37,787 INFO Set MAIN_POWER to False
2026-02-02 14:11:37,787 INFO [MOCK] Power OFF
2026-02-02 14:11:37,787 INFO [MOCK] 1553 Interface stopped
2026-02-02 14:11:42,794 INFO Check that MAIN_POWER is equal to 0
2026-02-02 14:11:42,794 INFO OK: read value = False
2026-02-02 14:11:46,797 INFO The test operator is required to switch on the target.
2026-02-02 14:11:46,797 INFO Set MAIN_POWER to True
2026-02-02 14:11:46,797 INFO [MOCK] Power ON
2026-02-02 14:11:46,797 INFO
====================================================================================================
[CRITICAL DEBUG] _initialize_field_values() CALLED
====================================================================================================
2026-02-02 14:11:46,797 INFO run_count = 1
2026-02-02 14:11:46,797 INFO SIMULATION_SCENARIO = 'mixed'
2026-02-02 14:11:46,797 INFO current_scenario = 'mixed'
2026-02-02 14:11:46,797 INFO _scenario_list = ['recycle_recovery_2', 'recycle_recovery_1', 'normal']
2026-02-02 14:11:46,797 INFO [DEBUG] Entering MIXED MODE branch
2026-02-02 14:11:46,797 INFO [DEBUG] Mixed mode: scenario_index=0, selected scenario='recycle_recovery_2'
2026-02-02 14:11:46,804 INFO [DEBUG] About to call _log_scenario_info('recycle_recovery_2', 1, 3)
2026-02-02 14:11:46,806 INFO
2026-02-02 14:11:46,806 INFO ####################################################################################################
2026-02-02 14:11:46,806 INFO ####################################################################################################
2026-02-02 14:11:46,806 INFO ### RUN 1/3: RECYCLE SELF-HEALING (2 CYCLES)
2026-02-02 14:11:46,806 INFO ####################################################################################################
2026-02-02 14:11:46,806 INFO
2026-02-02 14:11:46,807 INFO SCENARIO: recycle_recovery_2
2026-02-02 14:11:46,807 INFO DESCRIPTION: Processor fails, then auto-recovers after 2 RECYCLE events
2026-02-02 14:11:46,807 INFO
2026-02-02 14:11:46,807 INFO EXPECTED FAILURES:
2026-02-02 14:11:46,807 INFO ✗ Initial: processor_status = FAIL
2026-02-02 14:11:46,807 INFO ✗ Serial: 2x RECYCLE events detected
2026-02-02 14:11:46,808 INFO
2026-02-02 14:11:46,808 INFO EXPECTED PASSES:
2026-02-02 14:11:46,808 INFO ✓ After 2nd RECYCLE: processor_status = PASS
2026-02-02 14:11:46,809 INFO ✓ Final: All systems OK
2026-02-02 14:11:46,809 INFO
2026-02-02 14:11:46,809 INFO NOTES: Simulates persistent failure requiring multiple recovery attempts
2026-02-02 14:11:46,809 INFO
2026-02-02 14:11:46,810 INFO ####################################################################################################
2026-02-02 14:11:46,810 INFO ####################################################################################################
2026-02-02 14:11:46,810 INFO
2026-02-02 14:11:46,810 INFO [DEBUG] _log_scenario_info() completed
2026-02-02 14:11:46,812 INFO [MOCK APPLY] ⚠ RECYCLE scenario enabled: 2 cycles to recovery
2026-02-02 14:11:46,815 INFO [MOCK RECYCLE] Initial processor failure detected
2026-02-02 14:11:47,329 INFO [MOCK] Simulated RECYCLE event: %%F-10747-SP*: ***** RECYCLE!******
2026-02-02 14:11:47,329 INFO [MOCK RECYCLE] RECYCLE 1/2
2026-02-02 14:11:47,845 INFO [MOCK] Simulated RECYCLE event: %%F-10747-SP*: ***** RECYCLE!******
2026-02-02 14:11:47,845 INFO [MOCK RECYCLE] RECYCLE 2/2
2026-02-02 14:11:48,155 INFO [MOCK RECYCLE] System recovered after 2 RECYCLEs (slower self-healing)
2026-02-02 14:11:48,155 INFO [MOCK] 1553 Interface started - simulating BIT execution (Run 1)
2026-02-02 14:11:53,157 INFO Check that MAIN_POWER is equal to 1
2026-02-02 14:11:53,157 INFO OK: read value = True
2026-02-02 14:11:53,263 INFO Set A1_MsgRdrSettingsAndParameters settings_RDROperationalSettings_rdr_symbology_intensity to 100
2026-02-02 14:11:53,473 INFO [MOCK] BIT report available after 6.7s
2026-02-02 14:11:53,505 INFO BIT report available after 0.2s
2026-02-02 14:11:53,505 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_array_status is equal to false
2026-02-02 14:11:53,505 INFO OK: read value = false
2026-02-02 14:11:53,506 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_pedestal_status is equal to false
2026-02-02 14:11:53,506 WARNING CHECK ERROR: read value = true
2026-02-02 14:11:53,506 ERROR INTERFACE ERROR: Expected false, got true
2026-02-02 14:11:53,506 WARNING Known failure (ignored): radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_pedestal_status
2026-02-02 14:11:53,507 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_pressurization_status is equal to false
2026-02-02 14:11:53,507 INFO OK: read value = false
2026-02-02 14:11:53,507 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_processor_over_temperature_alarm is equal to false
2026-02-02 14:11:53,507 INFO OK: read value = false
2026-02-02 14:11:53,507 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_processor_status is equal to false
2026-02-02 14:11:53,508 INFO OK: read value = false
2026-02-02 14:11:53,508 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_receiver_status is equal to false
2026-02-02 14:11:53,508 INFO OK: read value = false
2026-02-02 14:11:53,508 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_rx_front_end_status is equal to false
2026-02-02 14:11:53,508 INFO OK: read value = false
2026-02-02 14:11:53,509 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_servoloop_over_temperature_alarm is equal to false
2026-02-02 14:11:53,509 INFO OK: read value = false
2026-02-02 14:11:53,509 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_servoloop_status is equal to false
2026-02-02 14:11:53,509 INFO OK: read value = false
2026-02-02 14:11:53,510 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_trasmitter_over_temperature_alarm is equal to false
2026-02-02 14:11:53,510 INFO OK: read value = false
2026-02-02 14:11:53,510 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_trasmitter_status is equal to false
2026-02-02 14:11:53,510 INFO OK: read value = false
2026-02-02 14:11:53,510 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_radar_fail_status is equal to false
2026-02-02 14:11:53,510 WARNING CHECK ERROR: read value = RDR_OK
2026-02-02 14:11:53,510 ERROR INTERFACE ERROR: Expected false, got RDR_OK
2026-02-02 14:11:53,510 WARNING Radar fail status: Known failure (caused only by pedestal)
2026-02-02 14:11:53,510 INFO [Run 1] B6 LRU Status: 12 total, 10 pass, 0 fail, 2 known
2026-02-02 14:11:53,512 INFO Known failures (HW setup): 2
2026-02-02 14:11:53,512 INFO
Forced B8 drill-down (FORCE_B8_DRILL_DOWN=True): Verifying all 173 B8 diagnostic fields...
2026-02-02 14:11:53,512 INFO [FORCE_B8_DRILL_DOWN] Performing B8 drill-down despite only known failures
2026-02-02 14:11:53,512 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_bcn_fail is equal to false
2026-02-02 14:11:53,512 INFO OK: read value = false
2026-02-02 14:11:53,512 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_gm_rbm_sea1_ta_wa_fail is equal to false
2026-02-02 14:11:53,512 INFO OK: read value = false
2026-02-02 14:11:53,512 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_group1_fail is equal to false
2026-02-02 14:11:53,512 INFO OK: read value = false
2026-02-02 14:11:53,512 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_group2_fail is equal to false
2026-02-02 14:11:53,513 INFO OK: read value = false
2026-02-02 14:11:53,513 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_group3_fail is equal to false
2026-02-02 14:11:53,513 INFO OK: read value = false
2026-02-02 14:11:53,513 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_group4_fail is equal to false
2026-02-02 14:11:53,513 INFO OK: read value = false
2026-02-02 14:11:53,513 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_group5_fail is equal to false
2026-02-02 14:11:53,513 INFO OK: read value = false
2026-02-02 14:11:53,513 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_hr_modes_and_gm_dbs_fail is equal to false
2026-02-02 14:11:53,515 INFO OK: read value = false
2026-02-02 14:11:53,515 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_no_rdr_symbology is equal to false
2026-02-02 14:11:53,515 INFO OK: read value = false
2026-02-02 14:11:53,515 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_not_identified_rdr_fail is equal to false
2026-02-02 14:11:53,515 WARNING CHECK ERROR: read value = 0
2026-02-02 14:11:53,515 ERROR INTERFACE ERROR: Expected false, got 0
2026-02-02 14:11:53,515 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_selected_channel_fail is equal to false
2026-02-02 14:11:53,515 INFO OK: read value = false
2026-02-02 14:11:53,515 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_total_rdr_fail is equal to false
2026-02-02 14:11:53,515 INFO OK: read value = false
2026-02-02 14:11:53,515 WARNING B8_Degradation: 1/12 failures
2026-02-02 14:11:53,515 INFO Check that B8_MsgBitReport failure_location_pedestal_FailureLocationPedestal_sru1_gimbal is equal to false
2026-02-02 14:11:53,515 INFO OK: read value = false
2026-02-02 14:11:53,515 INFO Check that B8_MsgBitReport failure_location_pedestal_FailureLocationPedestal_sru2_waveguide is equal to false
2026-02-02 14:11:53,515 INFO OK: read value = false
2026-02-02 14:11:53,515 INFO Check that B8_MsgBitReport failure_location_pedestal_FailureLocationPedestal_sru3_waveguide is equal to false
2026-02-02 14:11:53,515 INFO OK: read value = false
2026-02-02 14:11:53,515 INFO Check that B8_MsgBitReport failure_location_pedestal_FailureLocationPedestal_sru4_delta_guard_lna_switch is equal to false
2026-02-02 14:11:53,515 INFO OK: read value = false
2026-02-02 14:11:53,515 INFO Check that B8_MsgBitReport failure_location_pedestal_FailureLocationPedestal_sru5_waveguide_switch is equal to false
2026-02-02 14:11:53,515 INFO OK: read value = false
2026-02-02 14:11:53,515 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru10_main_computer is equal to false
2026-02-02 14:11:53,515 INFO OK: read value = false
2026-02-02 14:11:53,515 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru11_graphic_computer is equal to false
2026-02-02 14:11:53,515 INFO OK: read value = false
2026-02-02 14:11:53,515 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru12_power_supply is equal to false
2026-02-02 14:11:53,515 INFO OK: read value = false
2026-02-02 14:11:53,515 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru13_det_exp is equal to false
2026-02-02 14:11:53,515 INFO OK: read value = false
2026-02-02 14:11:53,515 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru14_rx_module is equal to false
2026-02-02 14:11:53,515 INFO OK: read value = false
2026-02-02 14:11:53,515 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru1_motherboard_chassis is equal to false
2026-02-02 14:11:53,515 INFO OK: read value = false
2026-02-02 14:11:53,515 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru2_mti_fft is equal to false
2026-02-02 14:11:53,515 INFO OK: read value = false
2026-02-02 14:11:53,515 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru3_dsp0 is equal to false
2026-02-02 14:11:53,515 INFO OK: read value = false
2026-02-02 14:11:53,520 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru4_dsp1 is equal to false
2026-02-02 14:11:53,520 INFO OK: read value = false
2026-02-02 14:11:53,520 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru5_cfar_px_ctrl is equal to false
2026-02-02 14:11:53,520 INFO OK: read value = false
2026-02-02 14:11:53,520 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru6_timer is equal to false
2026-02-02 14:11:53,520 INFO OK: read value = false
2026-02-02 14:11:53,521 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru7_post_processor is equal to false
2026-02-02 14:11:53,521 INFO OK: read value = false
2026-02-02 14:11:53,521 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru8_agc is equal to false
2026-02-02 14:11:53,523 INFO OK: read value = false
2026-02-02 14:11:53,523 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru9_esa_if is equal to false
2026-02-02 14:11:53,523 INFO OK: read value = false
2026-02-02 14:11:53,523 INFO Check that B8_MsgBitReport failure_location_receiver_FailureLocationReceiver_sru1_chassis is equal to false
2026-02-02 14:11:53,523 INFO OK: read value = false
2026-02-02 14:11:53,524 INFO Check that B8_MsgBitReport failure_location_receiver_FailureLocationReceiver_sru2_uhf_assy is equal to false
2026-02-02 14:11:53,524 INFO OK: read value = false
2026-02-02 14:11:53,524 INFO Check that B8_MsgBitReport failure_location_receiver_FailureLocationReceiver_sru3_synthesizer is equal to false
2026-02-02 14:11:53,524 INFO OK: read value = false
2026-02-02 14:11:53,525 INFO Check that B8_MsgBitReport failure_location_receiver_FailureLocationReceiver_sru4_delta_guard_down_converter is equal to false
2026-02-02 14:11:53,525 INFO OK: read value = false
2026-02-02 14:11:53,525 INFO Check that B8_MsgBitReport failure_location_receiver_FailureLocationReceiver_sru5_sum_down_converter is equal to false
2026-02-02 14:11:53,525 INFO OK: read value = false
2026-02-02 14:11:53,525 INFO Check that B8_MsgBitReport failure_location_receiver_FailureLocationReceiver_sru6_lo_distributor is equal to false
2026-02-02 14:11:53,525 INFO OK: read value = false
2026-02-02 14:11:53,525 INFO Check that B8_MsgBitReport failure_location_receiver_FailureLocationReceiver_sru7_up_converter is equal to false
2026-02-02 14:11:53,525 INFO OK: read value = false
2026-02-02 14:11:53,525 INFO Check that B8_MsgBitReport failure_location_rx_frontend_FailureLocationRxFrontEnd_sru1_chassis is equal to false
2026-02-02 14:11:53,525 INFO OK: read value = false
2026-02-02 14:11:53,526 INFO Check that B8_MsgBitReport failure_location_rx_frontend_FailureLocationRxFrontEnd_sru2_delta_guard_lna is equal to false
2026-02-02 14:11:53,526 INFO OK: read value = false
2026-02-02 14:11:53,526 INFO Check that B8_MsgBitReport failure_location_rx_frontend_FailureLocationRxFrontEnd_sru3_sum_act_prot_lna is equal to false
2026-02-02 14:11:53,526 INFO OK: read value = false
2026-02-02 14:11:53,527 INFO Check that B8_MsgBitReport failure_location_rx_frontend_FailureLocationRxFrontEnd_sru4_4port_circulator is equal to false
2026-02-02 14:11:53,527 INFO OK: read value = false
2026-02-02 14:11:53,527 INFO Check that B8_MsgBitReport failure_location_rx_frontend_FailureLocationRxFrontEnd_sru5_stc_delta_guard is equal to false
2026-02-02 14:11:53,527 INFO OK: read value = false
2026-02-02 14:11:53,527 INFO Check that B8_MsgBitReport failure_location_rx_frontend_FailureLocationRxFrontEnd_sru5_stc_sum is equal to false
2026-02-02 14:11:53,528 INFO OK: read value = false
2026-02-02 14:11:53,528 INFO Check that B8_MsgBitReport failure_location_servoloop_FailureLocationServoloop_sru1_chassis is equal to false
2026-02-02 14:11:53,529 INFO OK: read value = false
2026-02-02 14:11:53,529 INFO Check that B8_MsgBitReport failure_location_servoloop_FailureLocationServoloop_sru2_power_supply is equal to false
2026-02-02 14:11:53,529 INFO OK: read value = false
2026-02-02 14:11:53,529 INFO Check that B8_MsgBitReport failure_location_servoloop_FailureLocationServoloop_sru3_digital_controller is equal to false
2026-02-02 14:11:53,530 INFO OK: read value = false
2026-02-02 14:11:53,530 INFO Check that B8_MsgBitReport failure_location_transmitter_FailureLocationTransmitter_sru1_chassis is equal to false
2026-02-02 14:11:53,530 INFO OK: read value = false
2026-02-02 14:11:53,531 INFO Check that B8_MsgBitReport failure_location_transmitter_FailureLocationTransmitter_sru2_rex_f_tx is equal to false
2026-02-02 14:11:53,531 INFO OK: read value = false
2026-02-02 14:11:53,531 INFO Check that B8_MsgBitReport failure_location_transmitter_FailureLocationTransmitter_sru3_power_supply is equal to false
2026-02-02 14:11:53,531 INFO OK: read value = false
2026-02-02 14:11:53,531 INFO Check that B8_MsgBitReport failure_location_transmitter_FailureLocationTransmitter_sru4_valve_el_twt_tx is equal to false
2026-02-02 14:11:53,532 INFO OK: read value = false
2026-02-02 14:11:53,532 INFO Check that B8_MsgBitReport failure_location_transmitter_FailureLocationTransmitter_sru5_rf_driver is equal to false
2026-02-02 14:11:53,532 INFO OK: read value = false
2026-02-02 14:11:53,532 INFO Check that B8_MsgBitReport failure_location_transmitter_FailureLocationTransmitter_sru6_controller_tx is equal to false
2026-02-02 14:11:53,532 INFO OK: read value = false
2026-02-02 14:11:53,532 INFO Check that B8_MsgBitReport failure_location_transmitter_FailureLocationTransmitter_sru7_hv_power_supply is equal to false
2026-02-02 14:11:53,533 INFO OK: read value = false
2026-02-02 14:11:53,533 INFO Check that B8_MsgBitReport failure_location_transmitter_FailureLocationTransmitter_sru8_eht_power_supply is equal to false
2026-02-02 14:11:53,533 INFO OK: read value = false
2026-02-02 14:11:53,534 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc10_pulse_compressor_interface is equal to false
2026-02-02 14:11:53,534 INFO OK: read value = false
2026-02-02 14:11:53,534 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc11_dp_interface is equal to false
2026-02-02 14:11:53,534 INFO OK: read value = false
2026-02-02 14:11:53,534 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc13_taxi_running is equal to false
2026-02-02 14:11:53,535 INFO OK: read value = false
2026-02-02 14:11:53,535 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc14_external_xyp_ram is equal to false
2026-02-02 14:11:53,535 INFO OK: read value = false
2026-02-02 14:11:53,535 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc15_servoloop_interface is equal to false
2026-02-02 14:11:53,535 INFO OK: read value = false
2026-02-02 14:11:53,535 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc1_internal_xyp_ram is equal to false
2026-02-02 14:11:53,535 INFO OK: read value = false
2026-02-02 14:11:53,535 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc2_external_xyp_ram is equal to false
2026-02-02 14:11:53,535 INFO OK: read value = false
2026-02-02 14:11:53,535 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc5_dual_port_ram is equal to false
2026-02-02 14:11:53,535 INFO OK: read value = false
2026-02-02 14:11:53,535 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc6_agc_machine is equal to false
2026-02-02 14:11:53,535 INFO OK: read value = false
2026-02-02 14:11:53,535 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc7_sat_machine is equal to false
2026-02-02 14:11:53,535 INFO OK: read value = false
2026-02-02 14:11:53,535 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc9_c_ram_xy_checksum is equal to false
2026-02-02 14:11:53,535 INFO OK: read value = false
2026-02-02 14:11:53,535 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp10_video_memory is equal to false
2026-02-02 14:11:53,535 INFO OK: read value = false
2026-02-02 14:11:53,535 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp11_video_unit is equal to false
2026-02-02 14:11:53,535 INFO OK: read value = false
2026-02-02 14:11:53,535 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp12_transputer_unit is equal to false
2026-02-02 14:11:53,535 INFO OK: read value = false
2026-02-02 14:11:53,535 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp13_scan_converter_polar_memory is equal to false
2026-02-02 14:11:53,535 INFO OK: read value = false
2026-02-02 14:11:53,535 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp14_scan_converter_format_converter is equal to false
2026-02-02 14:11:53,535 INFO OK: read value = false
2026-02-02 14:11:53,535 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp1_486_cpu_tests is equal to false
2026-02-02 14:11:53,535 INFO OK: read value = false
2026-02-02 14:11:53,538 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp2_486_interfaces_with_r3000_gc is equal to false
2026-02-02 14:11:53,538 INFO OK: read value = false
2026-02-02 14:11:53,538 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp3_486_interface_with_slc is equal to false
2026-02-02 14:11:53,538 INFO OK: read value = false
2026-02-02 14:11:53,538 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp4_slc_communications is equal to false
2026-02-02 14:11:53,538 INFO OK: read value = false
2026-02-02 14:11:53,538 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp5_r3000_cpu_tests is equal to false
2026-02-02 14:11:53,538 INFO OK: read value = false
2026-02-02 14:11:53,540 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp6_r3000_interfaces is equal to false
2026-02-02 14:11:53,540 INFO OK: read value = false
2026-02-02 14:11:53,540 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp7_1553_and_discretes is equal to false
2026-02-02 14:11:53,540 INFO OK: read value = false
2026-02-02 14:11:53,540 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp8_graphic_cpu is equal to false
2026-02-02 14:11:53,540 INFO OK: read value = false
2026-02-02 14:11:53,541 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp9_graphic_processors is equal to false
2026-02-02 14:11:53,541 INFO OK: read value = false
2026-02-02 14:11:53,541 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_array_status is equal to false
2026-02-02 14:11:53,541 INFO OK: read value = false
2026-02-02 14:11:53,542 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_cal_delta_channel_fail is equal to false
2026-02-02 14:11:53,542 INFO OK: read value = false
2026-02-02 14:11:53,542 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_cal_injection_fail is equal to false
2026-02-02 14:11:53,542 INFO OK: read value = false
2026-02-02 14:11:53,543 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_cal_noise_fail is equal to false
2026-02-02 14:11:53,543 INFO OK: read value = false
2026-02-02 14:11:53,543 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_pedestal_status is equal to false
2026-02-02 14:11:53,543 INFO OK: read value = false
2026-02-02 14:11:53,543 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_processor_status is equal to false
2026-02-02 14:11:53,544 INFO OK: read value = false
2026-02-02 14:11:53,544 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_receiver_status is equal to false
2026-02-02 14:11:53,544 INFO OK: read value = false
2026-02-02 14:11:53,544 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_rx_frontend_status is equal to false
2026-02-02 14:11:53,544 INFO OK: read value = false
2026-02-02 14:11:53,544 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_servoloop_status is equal to false
2026-02-02 14:11:53,545 INFO OK: read value = false
2026-02-02 14:11:53,545 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_test_is1_upconverter_chain_levels is equal to false
2026-02-02 14:11:53,545 INFO OK: read value = false
2026-02-02 14:11:53,545 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_test_is2_downconverter_chain_levels is equal to false
2026-02-02 14:11:53,546 INFO OK: read value = false
2026-02-02 14:11:53,546 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_test_is3_antenna_status_inconsistent is equal to false
2026-02-02 14:11:53,546 INFO OK: read value = false
2026-02-02 14:11:53,547 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_test_is4_tx_status_inconsistent is equal to false
2026-02-02 14:11:53,547 INFO OK: read value = false
2026-02-02 14:11:53,547 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_test_is5_tx_power_level is equal to false
2026-02-02 14:11:53,547 INFO OK: read value = false
2026-02-02 14:11:53,548 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_transmitter_status is equal to false
2026-02-02 14:11:53,548 INFO OK: read value = false
2026-02-02 14:11:53,548 INFO Check that B8_MsgBitReport post_processor_test_results_PostProcessorTestResults_test_pp1_master_dsp is equal to false
2026-02-02 14:11:53,548 INFO OK: read value = false
2026-02-02 14:11:53,548 INFO Check that B8_MsgBitReport post_processor_test_results_PostProcessorTestResults_test_pp2_interface_card is equal to false
2026-02-02 14:11:53,549 INFO OK: read value = false
2026-02-02 14:11:53,549 INFO Check that B8_MsgBitReport post_processor_test_results_PostProcessorTestResults_test_pp3_cpu_cards is equal to false
2026-02-02 14:11:53,549 INFO OK: read value = false
2026-02-02 14:11:53,549 INFO Check that B8_MsgBitReport post_processor_test_results_PostProcessorTestResults_test_pp4_dma_bus is equal to false
2026-02-02 14:11:53,549 INFO OK: read value = false
2026-02-02 14:11:53,550 INFO Check that B8_MsgBitReport post_processor_test_results_PostProcessorTestResults_test_pp5_sp_interface is equal to false
2026-02-02 14:11:53,550 INFO OK: read value = false
2026-02-02 14:11:53,550 INFO Check that B8_MsgBitReport post_processor_test_results_PostProcessorTestResults_test_pp6_dp_interface is equal to false
2026-02-02 14:11:53,550 INFO OK: read value = false
2026-02-02 14:11:53,550 INFO Check that B8_MsgBitReport post_processor_test_results_PostProcessorTestResults_test_pp7_scan_converter_interface is equal to false
2026-02-02 14:11:53,551 INFO OK: read value = false
2026-02-02 14:11:53,551 INFO Check that B8_MsgBitReport post_processor_test_results_PostProcessorTestResults_test_pp8_agc_interface is equal to false
2026-02-02 14:11:53,551 INFO OK: read value = false
2026-02-02 14:11:53,551 INFO Check that B8_MsgBitReport power_supply_test_results_PowerSupplyTestResults_test_ps1_power_supply is equal to false
2026-02-02 14:11:53,551 INFO OK: read value = false
2026-02-02 14:11:53,551 INFO Check that B8_MsgBitReport power_supply_test_results_PowerSupplyTestResults_test_ps2_over_temperature is equal to false
2026-02-02 14:11:53,551 INFO OK: read value = false
2026-02-02 14:11:53,551 INFO Check that B8_MsgBitReport receiver_and_rx_frontend_test_results_ReceiverAndRxTestResults_test_fe1_lna is equal to false
2026-02-02 14:11:53,551 INFO OK: read value = false
2026-02-02 14:11:53,552 INFO Check that B8_MsgBitReport receiver_and_rx_frontend_test_results_ReceiverAndRxTestResults_test_fe2_agc_attenuators is equal to false
2026-02-02 14:11:53,552 INFO OK: read value = false
2026-02-02 14:11:53,552 INFO Check that B8_MsgBitReport receiver_and_rx_frontend_test_results_ReceiverAndRxTestResults_test_rx1_synthesizer_commands is equal to false
2026-02-02 14:11:53,552 INFO OK: read value = false
2026-02-02 14:11:53,552 INFO Check that B8_MsgBitReport receiver_and_rx_frontend_test_results_ReceiverAndRxTestResults_test_rx2_synthesizer_internal_tests is equal to false
2026-02-02 14:11:53,552 INFO OK: read value = false
2026-02-02 14:11:53,552 INFO Check that B8_MsgBitReport receiver_and_rx_frontend_test_results_ReceiverAndRxTestResults_test_rx3_uhf_oscillator_level is equal to false
2026-02-02 14:11:53,552 INFO OK: read value = false
2026-02-02 14:11:53,553 INFO Check that B8_MsgBitReport receiver_and_rx_frontend_test_results_ReceiverAndRxTestResults_test_rx4_downconverter_lo_level is equal to false
2026-02-02 14:11:53,553 INFO OK: read value = false
2026-02-02 14:11:53,553 INFO Check that B8_MsgBitReport receiver_and_rx_frontend_test_results_ReceiverAndRxTestResults_test_rx5_upconverter_lo_level is equal to false
2026-02-02 14:11:53,553 INFO OK: read value = false
2026-02-02 14:11:53,553 INFO Check that B8_MsgBitReport rx_module_test_results_RxModuleTestResults_test_rm16_calibration_sum_channel_fail is equal to false
2026-02-02 14:11:53,554 INFO OK: read value = false
2026-02-02 14:11:53,554 INFO Check that B8_MsgBitReport rx_module_test_results_RxModuleTestResults_test_rm1_master_clock_level is equal to false
2026-02-02 14:11:53,554 INFO OK: read value = false
2026-02-02 14:11:53,554 INFO Check that B8_MsgBitReport rx_module_test_results_RxModuleTestResults_test_rm2_expander_level is equal to false
2026-02-02 14:11:53,555 INFO OK: read value = false
2026-02-02 14:11:53,555 INFO Check that B8_MsgBitReport rx_module_test_results_RxModuleTestResults_test_rm3_sum_channel_down_converter is equal to false
2026-02-02 14:11:53,555 INFO OK: read value = false
2026-02-02 14:11:53,555 INFO Check that B8_MsgBitReport rx_module_test_results_RxModuleTestResults_test_rm4_dg_channel_down_converter is equal to false
2026-02-02 14:11:53,555 INFO OK: read value = false
2026-02-02 14:11:53,555 INFO Check that B8_MsgBitReport rx_module_test_results_RxModuleTestResults_test_rm5_noise_attenuators is equal to false
2026-02-02 14:11:53,556 INFO OK: read value = false
2026-02-02 14:11:53,556 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl10_agc_control is equal to false
2026-02-02 14:11:53,556 INFO OK: read value = false
2026-02-02 14:11:53,556 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl11_ad is equal to false
2026-02-02 14:11:53,556 INFO OK: read value = false
2026-02-02 14:11:53,556 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl12_das is equal to false
2026-02-02 14:11:53,556 INFO OK: read value = false
2026-02-02 14:11:53,557 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl13_serial_communications is equal to false
2026-02-02 14:11:53,557 INFO OK: read value = false
2026-02-02 14:11:53,557 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl14_taxi_interface is equal to false
2026-02-02 14:11:53,557 INFO OK: read value = false
2026-02-02 14:11:53,557 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl15_pedestal_centre_scan_location is equal to false
2026-02-02 14:11:53,558 INFO OK: read value = false
2026-02-02 14:11:53,558 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl1_low_voltage_power_supply is equal to false
2026-02-02 14:11:53,558 INFO OK: read value = false
2026-02-02 14:11:53,558 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl2_high_voltage_power_supply is equal to false
2026-02-02 14:11:53,558 INFO OK: read value = false
2026-02-02 14:11:53,558 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl3_motor_drivers is equal to false
2026-02-02 14:11:53,559 INFO OK: read value = false
2026-02-02 14:11:53,559 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl4_resolvers_power_supply is equal to false
2026-02-02 14:11:53,559 INFO OK: read value = false
2026-02-02 14:11:53,559 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl5_waveguide_switch is equal to false
2026-02-02 14:11:53,559 INFO OK: read value = false
2026-02-02 14:11:53,559 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl6_over_temperature is equal to false
2026-02-02 14:11:53,560 INFO OK: read value = false
2026-02-02 14:11:53,560 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl7_resolver_to_digital_conv is equal to false
2026-02-02 14:11:53,560 INFO OK: read value = false
2026-02-02 14:11:53,560 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl8_position_loop_error is equal to false
2026-02-02 14:11:53,560 INFO OK: read value = false
2026-02-02 14:11:53,561 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl9_microprocessor is equal to false
2026-02-02 14:11:53,561 INFO OK: read value = false
2026-02-02 14:11:53,561 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp10_board_overall is equal to false
2026-02-02 14:11:53,561 INFO OK: read value = false
2026-02-02 14:11:53,561 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp11_attenuatori_antenna is equal to false
2026-02-02 14:11:53,562 INFO OK: read value = false
2026-02-02 14:11:53,562 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp14_external_sp_if is equal to false
2026-02-02 14:11:53,562 INFO OK: read value = false
2026-02-02 14:11:53,562 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp16_bcn is equal to false
2026-02-02 14:11:53,562 INFO OK: read value = false
2026-02-02 14:11:53,563 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp1_timer1_up is equal to false
2026-02-02 14:11:53,563 INFO OK: read value = false
2026-02-02 14:11:53,563 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp2_timer_dma_pxc_if is equal to false
2026-02-02 14:11:53,563 INFO OK: read value = false
2026-02-02 14:11:53,563 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp3_timer_internal is equal to false
2026-02-02 14:11:53,563 INFO OK: read value = false
2026-02-02 14:11:53,563 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp4_px_ctrl_comm is equal to false
2026-02-02 14:11:53,564 INFO OK: read value = false
2026-02-02 14:11:53,564 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp5_video1_without_ad is equal to false
2026-02-02 14:11:53,564 INFO OK: read value = false
2026-02-02 14:11:53,564 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp6_video1_with_ad is equal to false
2026-02-02 14:11:53,564 INFO OK: read value = false
2026-02-02 14:11:53,565 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp7_video2_ad_sync is equal to false
2026-02-02 14:11:53,565 INFO OK: read value = false
2026-02-02 14:11:53,565 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp8_video2_timer_sync is equal to false
2026-02-02 14:11:53,565 INFO OK: read value = false
2026-02-02 14:11:53,565 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp9_ad_da is equal to false
2026-02-02 14:11:53,565 INFO OK: read value = false
2026-02-02 14:11:53,566 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp9b_wideband_expander is equal to false
2026-02-02 14:11:53,566 INFO OK: read value = false
2026-02-02 14:11:53,566 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx10_hv_ps_over_temperature_warning is equal to false
2026-02-02 14:11:53,566 INFO OK: read value = false
2026-02-02 14:11:53,566 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx11_twt_helix_over_current is equal to false
2026-02-02 14:11:53,567 INFO OK: read value = false
2026-02-02 14:11:53,567 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx12_cathode_to_helix_arc is equal to false
2026-02-02 14:11:53,567 INFO OK: read value = false
2026-02-02 14:11:53,567 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx13_twt_over_temperature_hazard is equal to false
2026-02-02 14:11:53,567 INFO OK: read value = false
2026-02-02 14:11:53,568 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx14_twt_over_temperature_warning is equal to false
2026-02-02 14:11:53,568 INFO OK: read value = false
2026-02-02 14:11:53,568 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx15_cathode_under_voltage is equal to false
2026-02-02 14:11:53,568 INFO OK: read value = false
2026-02-02 14:11:53,568 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx16_cathode_over_voltage is equal to false
2026-02-02 14:11:53,568 INFO OK: read value = false
2026-02-02 14:11:53,569 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx1_microprocessors is equal to false
2026-02-02 14:11:53,569 INFO OK: read value = false
2026-02-02 14:11:53,569 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx2_tx_rf_input is equal to false
2026-02-02 14:11:53,569 INFO OK: read value = false
2026-02-02 14:11:53,570 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx3_twt_rf_input is equal to false
2026-02-02 14:11:53,570 INFO OK: read value = false
2026-02-02 14:11:53,570 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx4_twt_rf_output is equal to false
2026-02-02 14:11:53,570 INFO OK: read value = false
2026-02-02 14:11:53,570 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx5_tx_rf_output_level is equal to false
2026-02-02 14:11:53,570 INFO OK: read value = false
2026-02-02 14:11:53,571 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx6_vswr is equal to false
2026-02-02 14:11:53,571 INFO OK: read value = false
2026-02-02 14:11:53,571 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx7_three_phase_input_power is equal to false
2026-02-02 14:11:53,571 INFO OK: read value = false
2026-02-02 14:11:53,571 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx8_low_voltage_power_supplies is equal to false
2026-02-02 14:11:53,572 INFO OK: read value = false
2026-02-02 14:11:53,572 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx9_hv_ps_over_temperature_hazard is equal to false
2026-02-02 14:11:53,572 INFO OK: read value = false
2026-02-02 14:11:53,572 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx17_collector_under_voltage is equal to false
2026-02-02 14:11:53,572 INFO OK: read value = false
2026-02-02 14:11:53,572 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx18_collector_over_voltage is equal to false
2026-02-02 14:11:53,574 INFO OK: read value = false
2026-02-02 14:11:53,574 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx19_rectified_voltage is equal to false
2026-02-02 14:11:53,574 INFO OK: read value = false
2026-02-02 14:11:53,574 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx20_cathode_inv_current_fail is equal to false
2026-02-02 14:11:53,574 INFO OK: read value = false
2026-02-02 14:11:53,574 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx21_collector_inv_current_fail is equal to false
2026-02-02 14:11:53,575 INFO OK: read value = false
2026-02-02 14:11:53,575 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx22_waveguide_pressurization is equal to false
2026-02-02 14:11:53,575 INFO OK: read value = false
2026-02-02 14:11:53,575 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx23_grid_window_over_duty_alt is equal to false
2026-02-02 14:11:53,575 INFO OK: read value = false
2026-02-02 14:11:53,576 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx24_floating_deck_fail is equal to false
2026-02-02 14:11:53,576 INFO OK: read value = false
2026-02-02 14:11:53,576 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx25_floating_deck_ps_fail is equal to false
2026-02-02 14:11:53,576 INFO OK: read value = false
2026-02-02 14:11:53,576 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx26_grid_window_over_duty is equal to false
2026-02-02 14:11:53,576 INFO OK: read value = false
2026-02-02 14:11:53,577 INFO [Run 1] B8 Diagnostics: 173 checked, 172 pass, 1 fail
2026-02-02 14:11:53,577 WARNING B8_Degradation: 1 failures - fail
2026-02-02 14:11:53,577 INFO 0.0s remaining ...
2026-02-02 14:11:53,577 INFO [Run 1] Serial: 2 total, 0 errors, 2 fatal, 2 recycles
2026-02-02 14:11:53,578 WARNING Session: GRIFO_M_PBIT : Repetition 1 of 3
Total Step 192: Success 189 (98.44%), Failed 3 (1.56%),
2026-02-02 14:11:53,578 INFO Repetition 2 of 3
2026-02-02 14:11:53,578 DEBUG [MOCK] Serial statistics reset
2026-02-02 14:11:53,578 INFO The test operator is required to switch off the target and wait 3 seconds.
2026-02-02 14:11:54,579 INFO Set MAIN_POWER to False
2026-02-02 14:11:54,579 INFO [MOCK] Power OFF
2026-02-02 14:11:54,579 INFO [MOCK] 1553 Interface stopped
2026-02-02 14:11:59,582 INFO Check that MAIN_POWER is equal to 0
2026-02-02 14:11:59,582 INFO OK: read value = False
2026-02-02 14:12:03,589 INFO The test operator is required to switch on the target.
2026-02-02 14:12:03,589 INFO Set MAIN_POWER to True
2026-02-02 14:12:03,589 INFO [MOCK] Power ON
2026-02-02 14:12:03,589 INFO
====================================================================================================
[CRITICAL DEBUG] _initialize_field_values() CALLED
====================================================================================================
2026-02-02 14:12:03,594 INFO run_count = 2
2026-02-02 14:12:03,594 INFO SIMULATION_SCENARIO = 'mixed'
2026-02-02 14:12:03,594 INFO current_scenario = 'mixed'
2026-02-02 14:12:03,594 INFO _scenario_list = ['recycle_recovery_2', 'recycle_recovery_1', 'normal']
2026-02-02 14:12:03,597 INFO [DEBUG] Entering MIXED MODE branch
2026-02-02 14:12:03,598 INFO [DEBUG] Mixed mode: scenario_index=1, selected scenario='recycle_recovery_1'
2026-02-02 14:12:03,599 INFO [DEBUG] About to call _log_scenario_info('recycle_recovery_1', 2, 3)
2026-02-02 14:12:03,599 INFO
2026-02-02 14:12:03,599 INFO ####################################################################################################
2026-02-02 14:12:03,599 INFO ####################################################################################################
2026-02-02 14:12:03,600 INFO ### RUN 2/3: RECYCLE SELF-HEALING (1 CYCLE)
2026-02-02 14:12:03,600 INFO ####################################################################################################
2026-02-02 14:12:03,600 INFO
2026-02-02 14:12:03,600 INFO SCENARIO: recycle_recovery_1
2026-02-02 14:12:03,600 INFO DESCRIPTION: Processor fails, then auto-recovers after 1 RECYCLE event
2026-02-02 14:12:03,602 INFO
2026-02-02 14:12:03,602 INFO EXPECTED FAILURES:
2026-02-02 14:12:03,602 INFO ✗ Initial: processor_status = FAIL
2026-02-02 14:12:03,602 INFO ✗ Serial: 1x RECYCLE event detected
2026-02-02 14:12:03,607 INFO
2026-02-02 14:12:03,607 INFO EXPECTED PASSES:
2026-02-02 14:12:03,608 INFO ✓ After RECYCLE: processor_status = PASS
2026-02-02 14:12:03,614 INFO ✓ Final: All systems OK
2026-02-02 14:12:03,614 INFO
2026-02-02 14:12:03,614 INFO NOTES: Simulates transient failure with fast self-healing. Test should detect RECYCLE and eventual recovery
2026-02-02 14:12:03,614 INFO
2026-02-02 14:12:03,616 INFO ####################################################################################################
2026-02-02 14:12:03,616 INFO ####################################################################################################
2026-02-02 14:12:03,616 INFO
2026-02-02 14:12:03,616 INFO [DEBUG] _log_scenario_info() completed
2026-02-02 14:12:03,616 INFO [MOCK APPLY] ⚠ RECYCLE scenario enabled: 1 cycle to recovery
2026-02-02 14:12:03,616 INFO [MOCK RECYCLE] Initial processor failure detected
2026-02-02 14:12:04,126 INFO [MOCK] Simulated RECYCLE event: %%F-10747-SP*: ***** RECYCLE!******
2026-02-02 14:12:04,438 INFO [MOCK RECYCLE] System recovered after 1 RECYCLE (self-healing)
2026-02-02 14:12:04,438 INFO [MOCK] 1553 Interface started - simulating BIT execution (Run 2)
2026-02-02 14:12:09,449 INFO Check that MAIN_POWER is equal to 1
2026-02-02 14:12:09,449 INFO OK: read value = True
2026-02-02 14:12:09,560 INFO Set A1_MsgRdrSettingsAndParameters settings_RDROperationalSettings_rdr_symbology_intensity to 100
2026-02-02 14:12:09,770 INFO [MOCK] BIT report available after 6.2s
2026-02-02 14:12:09,801 INFO BIT report available after 0.2s
2026-02-02 14:12:09,801 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_array_status is equal to false
2026-02-02 14:12:09,801 INFO OK: read value = false
2026-02-02 14:12:09,802 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_pedestal_status is equal to false
2026-02-02 14:12:09,803 WARNING CHECK ERROR: read value = true
2026-02-02 14:12:09,803 ERROR INTERFACE ERROR: Expected false, got true
2026-02-02 14:12:09,803 WARNING Known failure (ignored): radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_pedestal_status
2026-02-02 14:12:09,803 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_pressurization_status is equal to false
2026-02-02 14:12:09,803 INFO OK: read value = false
2026-02-02 14:12:09,803 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_processor_over_temperature_alarm is equal to false
2026-02-02 14:12:09,803 INFO OK: read value = false
2026-02-02 14:12:09,804 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_processor_status is equal to false
2026-02-02 14:12:09,804 INFO OK: read value = false
2026-02-02 14:12:09,804 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_receiver_status is equal to false
2026-02-02 14:12:09,804 INFO OK: read value = false
2026-02-02 14:12:09,804 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_rx_front_end_status is equal to false
2026-02-02 14:12:09,804 INFO OK: read value = false
2026-02-02 14:12:09,805 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_servoloop_over_temperature_alarm is equal to false
2026-02-02 14:12:09,805 INFO OK: read value = false
2026-02-02 14:12:09,805 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_servoloop_status is equal to false
2026-02-02 14:12:09,805 INFO OK: read value = false
2026-02-02 14:12:09,805 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_trasmitter_over_temperature_alarm is equal to false
2026-02-02 14:12:09,805 INFO OK: read value = false
2026-02-02 14:12:09,805 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_trasmitter_status is equal to false
2026-02-02 14:12:09,805 INFO OK: read value = false
2026-02-02 14:12:09,806 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_radar_fail_status is equal to false
2026-02-02 14:12:09,806 WARNING CHECK ERROR: read value = RDR_OK
2026-02-02 14:12:09,806 ERROR INTERFACE ERROR: Expected false, got RDR_OK
2026-02-02 14:12:09,807 WARNING Radar fail status: Known failure (caused only by pedestal)
2026-02-02 14:12:09,807 INFO [Run 2] B6 LRU Status: 12 total, 10 pass, 0 fail, 2 known
2026-02-02 14:12:09,807 INFO Known failures (HW setup): 2
2026-02-02 14:12:09,807 INFO
Forced B8 drill-down (FORCE_B8_DRILL_DOWN=True): Verifying all 173 B8 diagnostic fields...
2026-02-02 14:12:09,807 INFO [FORCE_B8_DRILL_DOWN] Performing B8 drill-down despite only known failures
2026-02-02 14:12:09,807 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_bcn_fail is equal to false
2026-02-02 14:12:09,807 INFO OK: read value = false
2026-02-02 14:12:09,808 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_gm_rbm_sea1_ta_wa_fail is equal to false
2026-02-02 14:12:09,808 INFO OK: read value = false
2026-02-02 14:12:09,808 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_group1_fail is equal to false
2026-02-02 14:12:09,808 INFO OK: read value = false
2026-02-02 14:12:09,808 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_group2_fail is equal to false
2026-02-02 14:12:09,809 INFO OK: read value = false
2026-02-02 14:12:09,809 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_group3_fail is equal to false
2026-02-02 14:12:09,809 INFO OK: read value = false
2026-02-02 14:12:09,810 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_group4_fail is equal to false
2026-02-02 14:12:09,810 INFO OK: read value = false
2026-02-02 14:12:09,810 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_group5_fail is equal to false
2026-02-02 14:12:09,810 INFO OK: read value = false
2026-02-02 14:12:09,810 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_hr_modes_and_gm_dbs_fail is equal to false
2026-02-02 14:12:09,810 INFO OK: read value = false
2026-02-02 14:12:09,810 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_no_rdr_symbology is equal to false
2026-02-02 14:12:09,811 INFO OK: read value = false
2026-02-02 14:12:09,811 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_not_identified_rdr_fail is equal to false
2026-02-02 14:12:09,811 WARNING CHECK ERROR: read value = 0
2026-02-02 14:12:09,811 ERROR INTERFACE ERROR: Expected false, got 0
2026-02-02 14:12:09,811 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_selected_channel_fail is equal to false
2026-02-02 14:12:09,811 INFO OK: read value = false
2026-02-02 14:12:09,812 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_total_rdr_fail is equal to false
2026-02-02 14:12:09,812 INFO OK: read value = false
2026-02-02 14:12:09,812 WARNING B8_Degradation: 1/12 failures
2026-02-02 14:12:09,812 INFO Check that B8_MsgBitReport failure_location_pedestal_FailureLocationPedestal_sru1_gimbal is equal to false
2026-02-02 14:12:09,812 INFO OK: read value = false
2026-02-02 14:12:09,813 INFO Check that B8_MsgBitReport failure_location_pedestal_FailureLocationPedestal_sru2_waveguide is equal to false
2026-02-02 14:12:09,813 INFO OK: read value = false
2026-02-02 14:12:09,813 INFO Check that B8_MsgBitReport failure_location_pedestal_FailureLocationPedestal_sru3_waveguide is equal to false
2026-02-02 14:12:09,813 INFO OK: read value = false
2026-02-02 14:12:09,813 INFO Check that B8_MsgBitReport failure_location_pedestal_FailureLocationPedestal_sru4_delta_guard_lna_switch is equal to false
2026-02-02 14:12:09,813 INFO OK: read value = false
2026-02-02 14:12:09,814 INFO Check that B8_MsgBitReport failure_location_pedestal_FailureLocationPedestal_sru5_waveguide_switch is equal to false
2026-02-02 14:12:09,814 INFO OK: read value = false
2026-02-02 14:12:09,814 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru10_main_computer is equal to false
2026-02-02 14:12:09,814 INFO OK: read value = false
2026-02-02 14:12:09,814 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru11_graphic_computer is equal to false
2026-02-02 14:12:09,814 INFO OK: read value = false
2026-02-02 14:12:09,815 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru12_power_supply is equal to false
2026-02-02 14:12:09,815 INFO OK: read value = false
2026-02-02 14:12:09,815 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru13_det_exp is equal to false
2026-02-02 14:12:09,816 INFO OK: read value = false
2026-02-02 14:12:09,816 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru14_rx_module is equal to false
2026-02-02 14:12:09,816 INFO OK: read value = false
2026-02-02 14:12:09,817 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru1_motherboard_chassis is equal to false
2026-02-02 14:12:09,817 INFO OK: read value = false
2026-02-02 14:12:09,817 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru2_mti_fft is equal to false
2026-02-02 14:12:09,817 INFO OK: read value = false
2026-02-02 14:12:09,817 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru3_dsp0 is equal to false
2026-02-02 14:12:09,817 INFO OK: read value = false
2026-02-02 14:12:09,817 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru4_dsp1 is equal to false
2026-02-02 14:12:09,817 INFO OK: read value = false
2026-02-02 14:12:09,817 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru5_cfar_px_ctrl is equal to false
2026-02-02 14:12:09,818 INFO OK: read value = false
2026-02-02 14:12:09,818 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru6_timer is equal to false
2026-02-02 14:12:09,818 INFO OK: read value = false
2026-02-02 14:12:09,818 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru7_post_processor is equal to false
2026-02-02 14:12:09,818 INFO OK: read value = false
2026-02-02 14:12:09,818 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru8_agc is equal to false
2026-02-02 14:12:09,818 INFO OK: read value = false
2026-02-02 14:12:09,818 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru9_esa_if is equal to false
2026-02-02 14:12:09,818 INFO OK: read value = false
2026-02-02 14:12:09,818 INFO Check that B8_MsgBitReport failure_location_receiver_FailureLocationReceiver_sru1_chassis is equal to false
2026-02-02 14:12:09,818 INFO OK: read value = false
2026-02-02 14:12:09,819 INFO Check that B8_MsgBitReport failure_location_receiver_FailureLocationReceiver_sru2_uhf_assy is equal to false
2026-02-02 14:12:09,819 INFO OK: read value = false
2026-02-02 14:12:09,819 INFO Check that B8_MsgBitReport failure_location_receiver_FailureLocationReceiver_sru3_synthesizer is equal to false
2026-02-02 14:12:09,819 INFO OK: read value = false
2026-02-02 14:12:09,819 INFO Check that B8_MsgBitReport failure_location_receiver_FailureLocationReceiver_sru4_delta_guard_down_converter is equal to false
2026-02-02 14:12:09,819 INFO OK: read value = false
2026-02-02 14:12:09,819 INFO Check that B8_MsgBitReport failure_location_receiver_FailureLocationReceiver_sru5_sum_down_converter is equal to false
2026-02-02 14:12:09,819 INFO OK: read value = false
2026-02-02 14:12:09,819 INFO Check that B8_MsgBitReport failure_location_receiver_FailureLocationReceiver_sru6_lo_distributor is equal to false
2026-02-02 14:12:09,819 INFO OK: read value = false
2026-02-02 14:12:09,819 INFO Check that B8_MsgBitReport failure_location_receiver_FailureLocationReceiver_sru7_up_converter is equal to false
2026-02-02 14:12:09,819 INFO OK: read value = false
2026-02-02 14:12:09,820 INFO Check that B8_MsgBitReport failure_location_rx_frontend_FailureLocationRxFrontEnd_sru1_chassis is equal to false
2026-02-02 14:12:09,820 INFO OK: read value = false
2026-02-02 14:12:09,821 INFO Check that B8_MsgBitReport failure_location_rx_frontend_FailureLocationRxFrontEnd_sru2_delta_guard_lna is equal to false
2026-02-02 14:12:09,821 INFO OK: read value = false
2026-02-02 14:12:09,821 INFO Check that B8_MsgBitReport failure_location_rx_frontend_FailureLocationRxFrontEnd_sru3_sum_act_prot_lna is equal to false
2026-02-02 14:12:09,821 INFO OK: read value = false
2026-02-02 14:12:09,823 INFO Check that B8_MsgBitReport failure_location_rx_frontend_FailureLocationRxFrontEnd_sru4_4port_circulator is equal to false
2026-02-02 14:12:09,823 INFO OK: read value = false
2026-02-02 14:12:09,824 INFO Check that B8_MsgBitReport failure_location_rx_frontend_FailureLocationRxFrontEnd_sru5_stc_delta_guard is equal to false
2026-02-02 14:12:09,824 INFO OK: read value = false
2026-02-02 14:12:09,825 INFO Check that B8_MsgBitReport failure_location_rx_frontend_FailureLocationRxFrontEnd_sru5_stc_sum is equal to false
2026-02-02 14:12:09,825 INFO OK: read value = false
2026-02-02 14:12:09,826 INFO Check that B8_MsgBitReport failure_location_servoloop_FailureLocationServoloop_sru1_chassis is equal to false
2026-02-02 14:12:09,826 INFO OK: read value = false
2026-02-02 14:12:09,826 INFO Check that B8_MsgBitReport failure_location_servoloop_FailureLocationServoloop_sru2_power_supply is equal to false
2026-02-02 14:12:09,827 INFO OK: read value = false
2026-02-02 14:12:09,827 INFO Check that B8_MsgBitReport failure_location_servoloop_FailureLocationServoloop_sru3_digital_controller is equal to false
2026-02-02 14:12:09,828 INFO OK: read value = false
2026-02-02 14:12:09,828 INFO Check that B8_MsgBitReport failure_location_transmitter_FailureLocationTransmitter_sru1_chassis is equal to false
2026-02-02 14:12:09,828 INFO OK: read value = false
2026-02-02 14:12:09,829 INFO Check that B8_MsgBitReport failure_location_transmitter_FailureLocationTransmitter_sru2_rex_f_tx is equal to false
2026-02-02 14:12:09,829 INFO OK: read value = false
2026-02-02 14:12:09,829 INFO Check that B8_MsgBitReport failure_location_transmitter_FailureLocationTransmitter_sru3_power_supply is equal to false
2026-02-02 14:12:09,830 INFO OK: read value = false
2026-02-02 14:12:09,830 INFO Check that B8_MsgBitReport failure_location_transmitter_FailureLocationTransmitter_sru4_valve_el_twt_tx is equal to false
2026-02-02 14:12:09,830 INFO OK: read value = false
2026-02-02 14:12:09,831 INFO Check that B8_MsgBitReport failure_location_transmitter_FailureLocationTransmitter_sru5_rf_driver is equal to false
2026-02-02 14:12:09,831 INFO OK: read value = false
2026-02-02 14:12:09,831 INFO Check that B8_MsgBitReport failure_location_transmitter_FailureLocationTransmitter_sru6_controller_tx is equal to false
2026-02-02 14:12:09,831 INFO OK: read value = false
2026-02-02 14:12:09,831 INFO Check that B8_MsgBitReport failure_location_transmitter_FailureLocationTransmitter_sru7_hv_power_supply is equal to false
2026-02-02 14:12:09,831 INFO OK: read value = false
2026-02-02 14:12:09,832 INFO Check that B8_MsgBitReport failure_location_transmitter_FailureLocationTransmitter_sru8_eht_power_supply is equal to false
2026-02-02 14:12:09,832 INFO OK: read value = false
2026-02-02 14:12:09,832 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc10_pulse_compressor_interface is equal to false
2026-02-02 14:12:09,832 INFO OK: read value = false
2026-02-02 14:12:09,832 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc11_dp_interface is equal to false
2026-02-02 14:12:09,832 INFO OK: read value = false
2026-02-02 14:12:09,833 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc13_taxi_running is equal to false
2026-02-02 14:12:09,833 INFO OK: read value = false
2026-02-02 14:12:09,833 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc14_external_xyp_ram is equal to false
2026-02-02 14:12:09,833 INFO OK: read value = false
2026-02-02 14:12:09,833 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc15_servoloop_interface is equal to false
2026-02-02 14:12:09,833 INFO OK: read value = false
2026-02-02 14:12:09,834 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc1_internal_xyp_ram is equal to false
2026-02-02 14:12:09,834 INFO OK: read value = false
2026-02-02 14:12:09,834 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc2_external_xyp_ram is equal to false
2026-02-02 14:12:09,834 INFO OK: read value = false
2026-02-02 14:12:09,835 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc5_dual_port_ram is equal to false
2026-02-02 14:12:09,835 INFO OK: read value = false
2026-02-02 14:12:09,835 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc6_agc_machine is equal to false
2026-02-02 14:12:09,835 INFO OK: read value = false
2026-02-02 14:12:09,835 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc7_sat_machine is equal to false
2026-02-02 14:12:09,836 INFO OK: read value = false
2026-02-02 14:12:09,836 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc9_c_ram_xy_checksum is equal to false
2026-02-02 14:12:09,836 INFO OK: read value = false
2026-02-02 14:12:09,836 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp10_video_memory is equal to false
2026-02-02 14:12:09,837 INFO OK: read value = false
2026-02-02 14:12:09,837 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp11_video_unit is equal to false
2026-02-02 14:12:09,837 INFO OK: read value = false
2026-02-02 14:12:09,837 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp12_transputer_unit is equal to false
2026-02-02 14:12:09,838 INFO OK: read value = false
2026-02-02 14:12:09,838 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp13_scan_converter_polar_memory is equal to false
2026-02-02 14:12:09,838 INFO OK: read value = false
2026-02-02 14:12:09,838 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp14_scan_converter_format_converter is equal to false
2026-02-02 14:12:09,838 INFO OK: read value = false
2026-02-02 14:12:09,839 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp1_486_cpu_tests is equal to false
2026-02-02 14:12:09,839 INFO OK: read value = false
2026-02-02 14:12:09,839 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp2_486_interfaces_with_r3000_gc is equal to false
2026-02-02 14:12:09,839 INFO OK: read value = false
2026-02-02 14:12:09,839 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp3_486_interface_with_slc is equal to false
2026-02-02 14:12:09,840 INFO OK: read value = false
2026-02-02 14:12:09,840 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp4_slc_communications is equal to false
2026-02-02 14:12:09,840 INFO OK: read value = false
2026-02-02 14:12:09,840 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp5_r3000_cpu_tests is equal to false
2026-02-02 14:12:09,840 INFO OK: read value = false
2026-02-02 14:12:09,840 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp6_r3000_interfaces is equal to false
2026-02-02 14:12:09,841 INFO OK: read value = false
2026-02-02 14:12:09,841 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp7_1553_and_discretes is equal to false
2026-02-02 14:12:09,841 INFO OK: read value = false
2026-02-02 14:12:09,841 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp8_graphic_cpu is equal to false
2026-02-02 14:12:09,841 INFO OK: read value = false
2026-02-02 14:12:09,841 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp9_graphic_processors is equal to false
2026-02-02 14:12:09,841 INFO OK: read value = false
2026-02-02 14:12:09,842 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_array_status is equal to false
2026-02-02 14:12:09,842 INFO OK: read value = false
2026-02-02 14:12:09,842 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_cal_delta_channel_fail is equal to false
2026-02-02 14:12:09,842 INFO OK: read value = false
2026-02-02 14:12:09,842 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_cal_injection_fail is equal to false
2026-02-02 14:12:09,843 INFO OK: read value = false
2026-02-02 14:12:09,843 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_cal_noise_fail is equal to false
2026-02-02 14:12:09,843 INFO OK: read value = false
2026-02-02 14:12:09,843 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_pedestal_status is equal to false
2026-02-02 14:12:09,843 INFO OK: read value = false
2026-02-02 14:12:09,843 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_processor_status is equal to false
2026-02-02 14:12:09,844 INFO OK: read value = false
2026-02-02 14:12:09,844 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_receiver_status is equal to false
2026-02-02 14:12:09,844 INFO OK: read value = false
2026-02-02 14:12:09,844 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_rx_frontend_status is equal to false
2026-02-02 14:12:09,845 INFO OK: read value = false
2026-02-02 14:12:09,845 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_servoloop_status is equal to false
2026-02-02 14:12:09,845 INFO OK: read value = false
2026-02-02 14:12:09,845 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_test_is1_upconverter_chain_levels is equal to false
2026-02-02 14:12:09,845 INFO OK: read value = false
2026-02-02 14:12:09,845 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_test_is2_downconverter_chain_levels is equal to false
2026-02-02 14:12:09,846 INFO OK: read value = false
2026-02-02 14:12:09,846 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_test_is3_antenna_status_inconsistent is equal to false
2026-02-02 14:12:09,846 INFO OK: read value = false
2026-02-02 14:12:09,846 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_test_is4_tx_status_inconsistent is equal to false
2026-02-02 14:12:09,846 INFO OK: read value = false
2026-02-02 14:12:09,847 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_test_is5_tx_power_level is equal to false
2026-02-02 14:12:09,847 INFO OK: read value = false
2026-02-02 14:12:09,847 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_transmitter_status is equal to false
2026-02-02 14:12:09,847 INFO OK: read value = false
2026-02-02 14:12:09,847 INFO Check that B8_MsgBitReport post_processor_test_results_PostProcessorTestResults_test_pp1_master_dsp is equal to false
2026-02-02 14:12:09,848 INFO OK: read value = false
2026-02-02 14:12:09,848 INFO Check that B8_MsgBitReport post_processor_test_results_PostProcessorTestResults_test_pp2_interface_card is equal to false
2026-02-02 14:12:09,848 INFO OK: read value = false
2026-02-02 14:12:09,848 INFO Check that B8_MsgBitReport post_processor_test_results_PostProcessorTestResults_test_pp3_cpu_cards is equal to false
2026-02-02 14:12:09,849 INFO OK: read value = false
2026-02-02 14:12:09,849 INFO Check that B8_MsgBitReport post_processor_test_results_PostProcessorTestResults_test_pp4_dma_bus is equal to false
2026-02-02 14:12:09,849 INFO OK: read value = false
2026-02-02 14:12:09,849 INFO Check that B8_MsgBitReport post_processor_test_results_PostProcessorTestResults_test_pp5_sp_interface is equal to false
2026-02-02 14:12:09,849 INFO OK: read value = false
2026-02-02 14:12:09,849 INFO Check that B8_MsgBitReport post_processor_test_results_PostProcessorTestResults_test_pp6_dp_interface is equal to false
2026-02-02 14:12:09,849 INFO OK: read value = false
2026-02-02 14:12:09,849 INFO Check that B8_MsgBitReport post_processor_test_results_PostProcessorTestResults_test_pp7_scan_converter_interface is equal to false
2026-02-02 14:12:09,849 INFO OK: read value = false
2026-02-02 14:12:09,850 INFO Check that B8_MsgBitReport post_processor_test_results_PostProcessorTestResults_test_pp8_agc_interface is equal to false
2026-02-02 14:12:09,850 INFO OK: read value = false
2026-02-02 14:12:09,850 INFO Check that B8_MsgBitReport power_supply_test_results_PowerSupplyTestResults_test_ps1_power_supply is equal to false
2026-02-02 14:12:09,850 INFO OK: read value = false
2026-02-02 14:12:09,850 INFO Check that B8_MsgBitReport power_supply_test_results_PowerSupplyTestResults_test_ps2_over_temperature is equal to false
2026-02-02 14:12:09,850 INFO OK: read value = false
2026-02-02 14:12:09,850 INFO Check that B8_MsgBitReport receiver_and_rx_frontend_test_results_ReceiverAndRxTestResults_test_fe1_lna is equal to false
2026-02-02 14:12:09,850 INFO OK: read value = false
2026-02-02 14:12:09,851 INFO Check that B8_MsgBitReport receiver_and_rx_frontend_test_results_ReceiverAndRxTestResults_test_fe2_agc_attenuators is equal to false
2026-02-02 14:12:09,851 INFO OK: read value = false
2026-02-02 14:12:09,851 INFO Check that B8_MsgBitReport receiver_and_rx_frontend_test_results_ReceiverAndRxTestResults_test_rx1_synthesizer_commands is equal to false
2026-02-02 14:12:09,851 INFO OK: read value = false
2026-02-02 14:12:09,851 INFO Check that B8_MsgBitReport receiver_and_rx_frontend_test_results_ReceiverAndRxTestResults_test_rx2_synthesizer_internal_tests is equal to false
2026-02-02 14:12:09,851 INFO OK: read value = false
2026-02-02 14:12:09,851 INFO Check that B8_MsgBitReport receiver_and_rx_frontend_test_results_ReceiverAndRxTestResults_test_rx3_uhf_oscillator_level is equal to false
2026-02-02 14:12:09,851 INFO OK: read value = false
2026-02-02 14:12:09,851 INFO Check that B8_MsgBitReport receiver_and_rx_frontend_test_results_ReceiverAndRxTestResults_test_rx4_downconverter_lo_level is equal to false
2026-02-02 14:12:09,851 INFO OK: read value = false
2026-02-02 14:12:09,851 INFO Check that B8_MsgBitReport receiver_and_rx_frontend_test_results_ReceiverAndRxTestResults_test_rx5_upconverter_lo_level is equal to false
2026-02-02 14:12:09,851 INFO OK: read value = false
2026-02-02 14:12:09,852 INFO Check that B8_MsgBitReport rx_module_test_results_RxModuleTestResults_test_rm16_calibration_sum_channel_fail is equal to false
2026-02-02 14:12:09,852 INFO OK: read value = false
2026-02-02 14:12:09,852 INFO Check that B8_MsgBitReport rx_module_test_results_RxModuleTestResults_test_rm1_master_clock_level is equal to false
2026-02-02 14:12:09,852 INFO OK: read value = false
2026-02-02 14:12:09,852 INFO Check that B8_MsgBitReport rx_module_test_results_RxModuleTestResults_test_rm2_expander_level is equal to false
2026-02-02 14:12:09,852 INFO OK: read value = false
2026-02-02 14:12:09,852 INFO Check that B8_MsgBitReport rx_module_test_results_RxModuleTestResults_test_rm3_sum_channel_down_converter is equal to false
2026-02-02 14:12:09,852 INFO OK: read value = false
2026-02-02 14:12:09,852 INFO Check that B8_MsgBitReport rx_module_test_results_RxModuleTestResults_test_rm4_dg_channel_down_converter is equal to false
2026-02-02 14:12:09,853 INFO OK: read value = false
2026-02-02 14:12:09,853 INFO Check that B8_MsgBitReport rx_module_test_results_RxModuleTestResults_test_rm5_noise_attenuators is equal to false
2026-02-02 14:12:09,853 INFO OK: read value = false
2026-02-02 14:12:09,853 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl10_agc_control is equal to false
2026-02-02 14:12:09,853 INFO OK: read value = false
2026-02-02 14:12:09,853 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl11_ad is equal to false
2026-02-02 14:12:09,853 INFO OK: read value = false
2026-02-02 14:12:09,853 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl12_das is equal to false
2026-02-02 14:12:09,853 INFO OK: read value = false
2026-02-02 14:12:09,854 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl13_serial_communications is equal to false
2026-02-02 14:12:09,854 INFO OK: read value = false
2026-02-02 14:12:09,854 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl14_taxi_interface is equal to false
2026-02-02 14:12:09,854 INFO OK: read value = false
2026-02-02 14:12:09,854 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl15_pedestal_centre_scan_location is equal to false
2026-02-02 14:12:09,856 INFO OK: read value = false
2026-02-02 14:12:09,856 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl1_low_voltage_power_supply is equal to false
2026-02-02 14:12:09,856 INFO OK: read value = false
2026-02-02 14:12:09,856 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl2_high_voltage_power_supply is equal to false
2026-02-02 14:12:09,856 INFO OK: read value = false
2026-02-02 14:12:09,857 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl3_motor_drivers is equal to false
2026-02-02 14:12:09,857 INFO OK: read value = false
2026-02-02 14:12:09,857 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl4_resolvers_power_supply is equal to false
2026-02-02 14:12:09,857 INFO OK: read value = false
2026-02-02 14:12:09,858 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl5_waveguide_switch is equal to false
2026-02-02 14:12:09,858 INFO OK: read value = false
2026-02-02 14:12:09,858 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl6_over_temperature is equal to false
2026-02-02 14:12:09,858 INFO OK: read value = false
2026-02-02 14:12:09,859 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl7_resolver_to_digital_conv is equal to false
2026-02-02 14:12:09,859 INFO OK: read value = false
2026-02-02 14:12:09,859 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl8_position_loop_error is equal to false
2026-02-02 14:12:09,859 INFO OK: read value = false
2026-02-02 14:12:09,859 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl9_microprocessor is equal to false
2026-02-02 14:12:09,859 INFO OK: read value = false
2026-02-02 14:12:09,860 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp10_board_overall is equal to false
2026-02-02 14:12:09,860 INFO OK: read value = false
2026-02-02 14:12:09,860 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp11_attenuatori_antenna is equal to false
2026-02-02 14:12:09,860 INFO OK: read value = false
2026-02-02 14:12:09,860 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp14_external_sp_if is equal to false
2026-02-02 14:12:09,861 INFO OK: read value = false
2026-02-02 14:12:09,861 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp16_bcn is equal to false
2026-02-02 14:12:09,861 INFO OK: read value = false
2026-02-02 14:12:09,861 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp1_timer1_up is equal to false
2026-02-02 14:12:09,861 INFO OK: read value = false
2026-02-02 14:12:09,861 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp2_timer_dma_pxc_if is equal to false
2026-02-02 14:12:09,861 INFO OK: read value = false
2026-02-02 14:12:09,862 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp3_timer_internal is equal to false
2026-02-02 14:12:09,862 INFO OK: read value = false
2026-02-02 14:12:09,862 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp4_px_ctrl_comm is equal to false
2026-02-02 14:12:09,862 INFO OK: read value = false
2026-02-02 14:12:09,863 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp5_video1_without_ad is equal to false
2026-02-02 14:12:09,863 INFO OK: read value = false
2026-02-02 14:12:09,863 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp6_video1_with_ad is equal to false
2026-02-02 14:12:09,863 INFO OK: read value = false
2026-02-02 14:12:09,863 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp7_video2_ad_sync is equal to false
2026-02-02 14:12:09,864 INFO OK: read value = false
2026-02-02 14:12:09,864 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp8_video2_timer_sync is equal to false
2026-02-02 14:12:09,864 INFO OK: read value = false
2026-02-02 14:12:09,865 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp9_ad_da is equal to false
2026-02-02 14:12:09,865 INFO OK: read value = false
2026-02-02 14:12:09,865 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp9b_wideband_expander is equal to false
2026-02-02 14:12:09,865 INFO OK: read value = false
2026-02-02 14:12:09,865 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx10_hv_ps_over_temperature_warning is equal to false
2026-02-02 14:12:09,866 INFO OK: read value = false
2026-02-02 14:12:09,866 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx11_twt_helix_over_current is equal to false
2026-02-02 14:12:09,866 INFO OK: read value = false
2026-02-02 14:12:09,866 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx12_cathode_to_helix_arc is equal to false
2026-02-02 14:12:09,866 INFO OK: read value = false
2026-02-02 14:12:09,866 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx13_twt_over_temperature_hazard is equal to false
2026-02-02 14:12:09,867 INFO OK: read value = false
2026-02-02 14:12:09,867 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx14_twt_over_temperature_warning is equal to false
2026-02-02 14:12:09,867 INFO OK: read value = false
2026-02-02 14:12:09,867 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx15_cathode_under_voltage is equal to false
2026-02-02 14:12:09,867 INFO OK: read value = false
2026-02-02 14:12:09,868 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx16_cathode_over_voltage is equal to false
2026-02-02 14:12:09,868 INFO OK: read value = false
2026-02-02 14:12:09,868 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx1_microprocessors is equal to false
2026-02-02 14:12:09,868 INFO OK: read value = false
2026-02-02 14:12:09,868 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx2_tx_rf_input is equal to false
2026-02-02 14:12:09,869 INFO OK: read value = false
2026-02-02 14:12:09,869 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx3_twt_rf_input is equal to false
2026-02-02 14:12:09,869 INFO OK: read value = false
2026-02-02 14:12:09,870 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx4_twt_rf_output is equal to false
2026-02-02 14:12:09,870 INFO OK: read value = false
2026-02-02 14:12:09,870 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx5_tx_rf_output_level is equal to false
2026-02-02 14:12:09,870 INFO OK: read value = false
2026-02-02 14:12:09,871 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx6_vswr is equal to false
2026-02-02 14:12:09,871 INFO OK: read value = false
2026-02-02 14:12:09,871 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx7_three_phase_input_power is equal to false
2026-02-02 14:12:09,871 INFO OK: read value = false
2026-02-02 14:12:09,871 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx8_low_voltage_power_supplies is equal to false
2026-02-02 14:12:09,872 INFO OK: read value = false
2026-02-02 14:12:09,873 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx9_hv_ps_over_temperature_hazard is equal to false
2026-02-02 14:12:09,873 INFO OK: read value = false
2026-02-02 14:12:09,873 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx17_collector_under_voltage is equal to false
2026-02-02 14:12:09,873 INFO OK: read value = false
2026-02-02 14:12:09,873 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx18_collector_over_voltage is equal to false
2026-02-02 14:12:09,874 INFO OK: read value = false
2026-02-02 14:12:09,874 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx19_rectified_voltage is equal to false
2026-02-02 14:12:09,874 INFO OK: read value = false
2026-02-02 14:12:09,874 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx20_cathode_inv_current_fail is equal to false
2026-02-02 14:12:09,874 INFO OK: read value = false
2026-02-02 14:12:09,875 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx21_collector_inv_current_fail is equal to false
2026-02-02 14:12:09,875 INFO OK: read value = false
2026-02-02 14:12:09,875 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx22_waveguide_pressurization is equal to false
2026-02-02 14:12:09,875 INFO OK: read value = false
2026-02-02 14:12:09,875 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx23_grid_window_over_duty_alt is equal to false
2026-02-02 14:12:09,876 INFO OK: read value = false
2026-02-02 14:12:09,876 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx24_floating_deck_fail is equal to false
2026-02-02 14:12:09,876 INFO OK: read value = false
2026-02-02 14:12:09,876 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx25_floating_deck_ps_fail is equal to false
2026-02-02 14:12:09,876 INFO OK: read value = false
2026-02-02 14:12:09,876 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx26_grid_window_over_duty is equal to false
2026-02-02 14:12:09,876 INFO OK: read value = false
2026-02-02 14:12:09,876 INFO [Run 2] B8 Diagnostics: 173 checked, 172 pass, 1 fail
2026-02-02 14:12:09,876 WARNING B8_Degradation: 1 failures - fail
2026-02-02 14:12:09,877 INFO 0.0s remaining ...
2026-02-02 14:12:09,877 INFO [Run 2] Serial: 1 total, 0 errors, 1 fatal, 1 recycles
2026-02-02 14:12:09,877 WARNING Session: GRIFO_M_PBIT : Repetition 2 of 3
Total Step 190: Success 187 (98.42%), Failed 3 (1.58%),
2026-02-02 14:12:09,877 INFO Repetition 3 of 3
2026-02-02 14:12:09,877 DEBUG [MOCK] Serial statistics reset
2026-02-02 14:12:09,877 INFO The test operator is required to switch off the target and wait 3 seconds.
2026-02-02 14:12:10,890 INFO Set MAIN_POWER to False
2026-02-02 14:12:10,890 INFO [MOCK] Power OFF
2026-02-02 14:12:10,890 INFO [MOCK] 1553 Interface stopped
2026-02-02 14:12:15,891 INFO Check that MAIN_POWER is equal to 0
2026-02-02 14:12:15,891 INFO OK: read value = False
2026-02-02 14:12:19,902 INFO The test operator is required to switch on the target.
2026-02-02 14:12:19,902 INFO Set MAIN_POWER to True
2026-02-02 14:12:19,902 INFO [MOCK] Power ON
2026-02-02 14:12:19,902 INFO
====================================================================================================
[CRITICAL DEBUG] _initialize_field_values() CALLED
====================================================================================================
2026-02-02 14:12:19,902 INFO run_count = 3
2026-02-02 14:12:19,902 INFO SIMULATION_SCENARIO = 'mixed'
2026-02-02 14:12:19,902 INFO current_scenario = 'mixed'
2026-02-02 14:12:19,902 INFO _scenario_list = ['recycle_recovery_2', 'recycle_recovery_1', 'normal']
2026-02-02 14:12:19,902 INFO [DEBUG] Entering MIXED MODE branch
2026-02-02 14:12:19,902 INFO [DEBUG] Mixed mode: scenario_index=2, selected scenario='normal'
2026-02-02 14:12:19,902 INFO [DEBUG] About to call _log_scenario_info('normal', 3, 3)
2026-02-02 14:12:19,902 INFO
2026-02-02 14:12:19,902 INFO ####################################################################################################
2026-02-02 14:12:19,902 INFO ####################################################################################################
2026-02-02 14:12:19,902 INFO ### RUN 3/3: NORMAL OPERATION
2026-02-02 14:12:19,902 INFO ####################################################################################################
2026-02-02 14:12:19,902 INFO
2026-02-02 14:12:19,902 INFO SCENARIO: normal
2026-02-02 14:12:19,902 INFO DESCRIPTION: All systems operational - no failures
2026-02-02 14:12:19,902 INFO
2026-02-02 14:12:19,902 INFO EXPECTED FAILURES:
2026-02-02 14:12:19,902 INFO (none)
2026-02-02 14:12:19,902 INFO
2026-02-02 14:12:19,902 INFO EXPECTED PASSES:
2026-02-02 14:12:19,902 INFO ✓ All B6 status fields (except pedestal)
2026-02-02 14:12:19,908 INFO ✓ All B8 diagnostic fields
2026-02-02 14:12:19,908 INFO ✓ Target generation
2026-02-02 14:12:19,908 INFO
2026-02-02 14:12:19,909 INFO NOTES: Pedestal always fails due to HW test setup limitation (unit not present)
2026-02-02 14:12:19,909 INFO
2026-02-02 14:12:19,909 INFO ####################################################################################################
2026-02-02 14:12:19,909 INFO ####################################################################################################
2026-02-02 14:12:19,909 INFO
2026-02-02 14:12:19,910 INFO [DEBUG] _log_scenario_info() completed
2026-02-02 14:12:19,910 INFO [MOCK APPLY] ✓ All systems nominal (no failures injected)
2026-02-02 14:12:19,910 INFO [MOCK RECYCLE] Initial processor failure detected
2026-02-02 14:12:20,421 INFO [MOCK] Simulated RECYCLE event: %%F-10747-SP*: ***** RECYCLE!******
2026-02-02 14:12:20,733 INFO [MOCK RECYCLE] System recovered after 1 RECYCLE (self-healing)
2026-02-02 14:12:20,733 INFO [MOCK] 1553 Interface started - simulating BIT execution (Run 3)
2026-02-02 14:12:25,741 INFO Check that MAIN_POWER is equal to 1
2026-02-02 14:12:25,741 INFO OK: read value = True
2026-02-02 14:12:25,852 INFO Set A1_MsgRdrSettingsAndParameters settings_RDROperationalSettings_rdr_symbology_intensity to 100
2026-02-02 14:12:26,063 INFO [MOCK] BIT report available after 6.2s
2026-02-02 14:12:26,093 INFO BIT report available after 0.2s
2026-02-02 14:12:26,094 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_array_status is equal to false
2026-02-02 14:12:26,094 INFO OK: read value = false
2026-02-02 14:12:26,094 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_pedestal_status is equal to false
2026-02-02 14:12:26,095 WARNING CHECK ERROR: read value = true
2026-02-02 14:12:26,095 ERROR INTERFACE ERROR: Expected false, got true
2026-02-02 14:12:26,095 WARNING Known failure (ignored): radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_pedestal_status
2026-02-02 14:12:26,096 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_pressurization_status is equal to false
2026-02-02 14:12:26,096 INFO OK: read value = false
2026-02-02 14:12:26,097 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_processor_over_temperature_alarm is equal to false
2026-02-02 14:12:26,097 INFO OK: read value = false
2026-02-02 14:12:26,097 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_processor_status is equal to false
2026-02-02 14:12:26,098 INFO OK: read value = false
2026-02-02 14:12:26,098 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_receiver_status is equal to false
2026-02-02 14:12:26,098 INFO OK: read value = false
2026-02-02 14:12:26,099 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_rx_front_end_status is equal to false
2026-02-02 14:12:26,100 INFO OK: read value = false
2026-02-02 14:12:26,100 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_servoloop_over_temperature_alarm is equal to false
2026-02-02 14:12:26,100 INFO OK: read value = false
2026-02-02 14:12:26,101 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_servoloop_status is equal to false
2026-02-02 14:12:26,101 INFO OK: read value = false
2026-02-02 14:12:26,101 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_trasmitter_over_temperature_alarm is equal to false
2026-02-02 14:12:26,101 INFO OK: read value = false
2026-02-02 14:12:26,101 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_trasmitter_status is equal to false
2026-02-02 14:12:26,101 INFO OK: read value = false
2026-02-02 14:12:26,102 INFO Check that B6_MsgRdrSettingsAndParametersTellback radar_health_status_and_bit_report_valid_RdrHealthStatusAndBitReport_radar_fail_status is equal to false
2026-02-02 14:12:26,102 WARNING CHECK ERROR: read value = RDR_OK
2026-02-02 14:12:26,102 ERROR INTERFACE ERROR: Expected false, got RDR_OK
2026-02-02 14:12:26,102 WARNING Radar fail status: Known failure (caused only by pedestal)
2026-02-02 14:12:26,102 INFO [Run 3] B6 LRU Status: 12 total, 10 pass, 0 fail, 2 known
2026-02-02 14:12:26,102 INFO Known failures (HW setup): 2
2026-02-02 14:12:26,103 INFO
Forced B8 drill-down (FORCE_B8_DRILL_DOWN=True): Verifying all 173 B8 diagnostic fields...
2026-02-02 14:12:26,103 INFO [FORCE_B8_DRILL_DOWN] Performing B8 drill-down despite only known failures
2026-02-02 14:12:26,103 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_bcn_fail is equal to false
2026-02-02 14:12:26,103 INFO OK: read value = false
2026-02-02 14:12:26,103 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_gm_rbm_sea1_ta_wa_fail is equal to false
2026-02-02 14:12:26,104 INFO OK: read value = false
2026-02-02 14:12:26,104 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_group1_fail is equal to false
2026-02-02 14:12:26,104 INFO OK: read value = false
2026-02-02 14:12:26,104 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_group2_fail is equal to false
2026-02-02 14:12:26,104 INFO OK: read value = false
2026-02-02 14:12:26,104 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_group3_fail is equal to false
2026-02-02 14:12:26,106 INFO OK: read value = false
2026-02-02 14:12:26,106 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_group4_fail is equal to false
2026-02-02 14:12:26,106 INFO OK: read value = false
2026-02-02 14:12:26,106 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_group5_fail is equal to false
2026-02-02 14:12:26,106 INFO OK: read value = false
2026-02-02 14:12:26,107 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_hr_modes_and_gm_dbs_fail is equal to false
2026-02-02 14:12:26,107 INFO OK: read value = false
2026-02-02 14:12:26,107 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_no_rdr_symbology is equal to false
2026-02-02 14:12:26,107 INFO OK: read value = false
2026-02-02 14:12:26,107 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_not_identified_rdr_fail is equal to false
2026-02-02 14:12:26,107 WARNING CHECK ERROR: read value = 0
2026-02-02 14:12:26,107 ERROR INTERFACE ERROR: Expected false, got 0
2026-02-02 14:12:26,108 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_selected_channel_fail is equal to false
2026-02-02 14:12:26,108 INFO OK: read value = false
2026-02-02 14:12:26,108 INFO Check that B8_MsgBitReport degradation_conditions_w1_DegradationConditionsW1_total_rdr_fail is equal to false
2026-02-02 14:12:26,111 INFO OK: read value = false
2026-02-02 14:12:26,111 WARNING B8_Degradation: 1/12 failures
2026-02-02 14:12:26,111 INFO Check that B8_MsgBitReport failure_location_pedestal_FailureLocationPedestal_sru1_gimbal is equal to false
2026-02-02 14:12:26,111 INFO OK: read value = false
2026-02-02 14:12:26,112 INFO Check that B8_MsgBitReport failure_location_pedestal_FailureLocationPedestal_sru2_waveguide is equal to false
2026-02-02 14:12:26,112 INFO OK: read value = false
2026-02-02 14:12:26,112 INFO Check that B8_MsgBitReport failure_location_pedestal_FailureLocationPedestal_sru3_waveguide is equal to false
2026-02-02 14:12:26,112 INFO OK: read value = false
2026-02-02 14:12:26,112 INFO Check that B8_MsgBitReport failure_location_pedestal_FailureLocationPedestal_sru4_delta_guard_lna_switch is equal to false
2026-02-02 14:12:26,113 INFO OK: read value = false
2026-02-02 14:12:26,113 INFO Check that B8_MsgBitReport failure_location_pedestal_FailureLocationPedestal_sru5_waveguide_switch is equal to false
2026-02-02 14:12:26,113 INFO OK: read value = false
2026-02-02 14:12:26,113 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru10_main_computer is equal to false
2026-02-02 14:12:26,113 INFO OK: read value = false
2026-02-02 14:12:26,113 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru11_graphic_computer is equal to false
2026-02-02 14:12:26,113 INFO OK: read value = false
2026-02-02 14:12:26,113 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru12_power_supply is equal to false
2026-02-02 14:12:26,113 INFO OK: read value = false
2026-02-02 14:12:26,113 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru13_det_exp is equal to false
2026-02-02 14:12:26,114 INFO OK: read value = false
2026-02-02 14:12:26,114 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru14_rx_module is equal to false
2026-02-02 14:12:26,114 INFO OK: read value = false
2026-02-02 14:12:26,114 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru1_motherboard_chassis is equal to false
2026-02-02 14:12:26,114 INFO OK: read value = false
2026-02-02 14:12:26,114 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru2_mti_fft is equal to false
2026-02-02 14:12:26,114 INFO OK: read value = false
2026-02-02 14:12:26,114 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru3_dsp0 is equal to false
2026-02-02 14:12:26,114 INFO OK: read value = false
2026-02-02 14:12:26,114 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru4_dsp1 is equal to false
2026-02-02 14:12:26,114 INFO OK: read value = false
2026-02-02 14:12:26,114 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru5_cfar_px_ctrl is equal to false
2026-02-02 14:12:26,115 INFO OK: read value = false
2026-02-02 14:12:26,115 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru6_timer is equal to false
2026-02-02 14:12:26,115 INFO OK: read value = false
2026-02-02 14:12:26,115 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru7_post_processor is equal to false
2026-02-02 14:12:26,115 INFO OK: read value = false
2026-02-02 14:12:26,115 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru8_agc is equal to false
2026-02-02 14:12:26,115 INFO OK: read value = false
2026-02-02 14:12:26,115 INFO Check that B8_MsgBitReport failure_location_processor_FailureLocationProcessor_sru9_esa_if is equal to false
2026-02-02 14:12:26,116 INFO OK: read value = false
2026-02-02 14:12:26,116 INFO Check that B8_MsgBitReport failure_location_receiver_FailureLocationReceiver_sru1_chassis is equal to false
2026-02-02 14:12:26,116 INFO OK: read value = false
2026-02-02 14:12:26,116 INFO Check that B8_MsgBitReport failure_location_receiver_FailureLocationReceiver_sru2_uhf_assy is equal to false
2026-02-02 14:12:26,116 INFO OK: read value = false
2026-02-02 14:12:26,116 INFO Check that B8_MsgBitReport failure_location_receiver_FailureLocationReceiver_sru3_synthesizer is equal to false
2026-02-02 14:12:26,116 INFO OK: read value = false
2026-02-02 14:12:26,116 INFO Check that B8_MsgBitReport failure_location_receiver_FailureLocationReceiver_sru4_delta_guard_down_converter is equal to false
2026-02-02 14:12:26,116 INFO OK: read value = false
2026-02-02 14:12:26,116 INFO Check that B8_MsgBitReport failure_location_receiver_FailureLocationReceiver_sru5_sum_down_converter is equal to false
2026-02-02 14:12:26,116 INFO OK: read value = false
2026-02-02 14:12:26,117 INFO Check that B8_MsgBitReport failure_location_receiver_FailureLocationReceiver_sru6_lo_distributor is equal to false
2026-02-02 14:12:26,117 INFO OK: read value = false
2026-02-02 14:12:26,117 INFO Check that B8_MsgBitReport failure_location_receiver_FailureLocationReceiver_sru7_up_converter is equal to false
2026-02-02 14:12:26,117 INFO OK: read value = false
2026-02-02 14:12:26,117 INFO Check that B8_MsgBitReport failure_location_rx_frontend_FailureLocationRxFrontEnd_sru1_chassis is equal to false
2026-02-02 14:12:26,117 INFO OK: read value = false
2026-02-02 14:12:26,117 INFO Check that B8_MsgBitReport failure_location_rx_frontend_FailureLocationRxFrontEnd_sru2_delta_guard_lna is equal to false
2026-02-02 14:12:26,117 INFO OK: read value = false
2026-02-02 14:12:26,117 INFO Check that B8_MsgBitReport failure_location_rx_frontend_FailureLocationRxFrontEnd_sru3_sum_act_prot_lna is equal to false
2026-02-02 14:12:26,117 INFO OK: read value = false
2026-02-02 14:12:26,117 INFO Check that B8_MsgBitReport failure_location_rx_frontend_FailureLocationRxFrontEnd_sru4_4port_circulator is equal to false
2026-02-02 14:12:26,118 INFO OK: read value = false
2026-02-02 14:12:26,118 INFO Check that B8_MsgBitReport failure_location_rx_frontend_FailureLocationRxFrontEnd_sru5_stc_delta_guard is equal to false
2026-02-02 14:12:26,118 INFO OK: read value = false
2026-02-02 14:12:26,118 INFO Check that B8_MsgBitReport failure_location_rx_frontend_FailureLocationRxFrontEnd_sru5_stc_sum is equal to false
2026-02-02 14:12:26,119 INFO OK: read value = false
2026-02-02 14:12:26,119 INFO Check that B8_MsgBitReport failure_location_servoloop_FailureLocationServoloop_sru1_chassis is equal to false
2026-02-02 14:12:26,119 INFO OK: read value = false
2026-02-02 14:12:26,119 INFO Check that B8_MsgBitReport failure_location_servoloop_FailureLocationServoloop_sru2_power_supply is equal to false
2026-02-02 14:12:26,120 INFO OK: read value = false
2026-02-02 14:12:26,120 INFO Check that B8_MsgBitReport failure_location_servoloop_FailureLocationServoloop_sru3_digital_controller is equal to false
2026-02-02 14:12:26,120 INFO OK: read value = false
2026-02-02 14:12:26,121 INFO Check that B8_MsgBitReport failure_location_transmitter_FailureLocationTransmitter_sru1_chassis is equal to false
2026-02-02 14:12:26,121 INFO OK: read value = false
2026-02-02 14:12:26,121 INFO Check that B8_MsgBitReport failure_location_transmitter_FailureLocationTransmitter_sru2_rex_f_tx is equal to false
2026-02-02 14:12:26,121 INFO OK: read value = false
2026-02-02 14:12:26,121 INFO Check that B8_MsgBitReport failure_location_transmitter_FailureLocationTransmitter_sru3_power_supply is equal to false
2026-02-02 14:12:26,121 INFO OK: read value = false
2026-02-02 14:12:26,121 INFO Check that B8_MsgBitReport failure_location_transmitter_FailureLocationTransmitter_sru4_valve_el_twt_tx is equal to false
2026-02-02 14:12:26,121 INFO OK: read value = false
2026-02-02 14:12:26,121 INFO Check that B8_MsgBitReport failure_location_transmitter_FailureLocationTransmitter_sru5_rf_driver is equal to false
2026-02-02 14:12:26,121 INFO OK: read value = false
2026-02-02 14:12:26,121 INFO Check that B8_MsgBitReport failure_location_transmitter_FailureLocationTransmitter_sru6_controller_tx is equal to false
2026-02-02 14:12:26,121 INFO OK: read value = false
2026-02-02 14:12:26,122 INFO Check that B8_MsgBitReport failure_location_transmitter_FailureLocationTransmitter_sru7_hv_power_supply is equal to false
2026-02-02 14:12:26,122 INFO OK: read value = false
2026-02-02 14:12:26,122 INFO Check that B8_MsgBitReport failure_location_transmitter_FailureLocationTransmitter_sru8_eht_power_supply is equal to false
2026-02-02 14:12:26,122 INFO OK: read value = false
2026-02-02 14:12:26,122 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc10_pulse_compressor_interface is equal to false
2026-02-02 14:12:26,122 INFO OK: read value = false
2026-02-02 14:12:26,122 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc11_dp_interface is equal to false
2026-02-02 14:12:26,123 INFO OK: read value = false
2026-02-02 14:12:26,123 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc13_taxi_running is equal to false
2026-02-02 14:12:26,123 INFO OK: read value = false
2026-02-02 14:12:26,123 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc14_external_xyp_ram is equal to false
2026-02-02 14:12:26,123 INFO OK: read value = false
2026-02-02 14:12:26,124 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc15_servoloop_interface is equal to false
2026-02-02 14:12:26,124 INFO OK: read value = false
2026-02-02 14:12:26,124 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc1_internal_xyp_ram is equal to false
2026-02-02 14:12:26,128 INFO OK: read value = false
2026-02-02 14:12:26,128 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc2_external_xyp_ram is equal to false
2026-02-02 14:12:26,128 INFO OK: read value = false
2026-02-02 14:12:26,128 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc5_dual_port_ram is equal to false
2026-02-02 14:12:26,128 INFO OK: read value = false
2026-02-02 14:12:26,128 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc6_agc_machine is equal to false
2026-02-02 14:12:26,128 INFO OK: read value = false
2026-02-02 14:12:26,128 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc7_sat_machine is equal to false
2026-02-02 14:12:26,128 INFO OK: read value = false
2026-02-02 14:12:26,130 INFO Check that B8_MsgBitReport agc_test_results_AGCTestResults_test_agc9_c_ram_xy_checksum is equal to false
2026-02-02 14:12:26,130 INFO OK: read value = false
2026-02-02 14:12:26,130 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp10_video_memory is equal to false
2026-02-02 14:12:26,130 INFO OK: read value = false
2026-02-02 14:12:26,130 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp11_video_unit is equal to false
2026-02-02 14:12:26,130 INFO OK: read value = false
2026-02-02 14:12:26,130 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp12_transputer_unit is equal to false
2026-02-02 14:12:26,130 INFO OK: read value = false
2026-02-02 14:12:26,130 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp13_scan_converter_polar_memory is equal to false
2026-02-02 14:12:26,130 INFO OK: read value = false
2026-02-02 14:12:26,130 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp14_scan_converter_format_converter is equal to false
2026-02-02 14:12:26,131 INFO OK: read value = false
2026-02-02 14:12:26,131 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp1_486_cpu_tests is equal to false
2026-02-02 14:12:26,131 INFO OK: read value = false
2026-02-02 14:12:26,131 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp2_486_interfaces_with_r3000_gc is equal to false
2026-02-02 14:12:26,131 INFO OK: read value = false
2026-02-02 14:12:26,131 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp3_486_interface_with_slc is equal to false
2026-02-02 14:12:26,131 INFO OK: read value = false
2026-02-02 14:12:26,131 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp4_slc_communications is equal to false
2026-02-02 14:12:26,131 INFO OK: read value = false
2026-02-02 14:12:26,131 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp5_r3000_cpu_tests is equal to false
2026-02-02 14:12:26,131 INFO OK: read value = false
2026-02-02 14:12:26,132 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp6_r3000_interfaces is equal to false
2026-02-02 14:12:26,132 INFO OK: read value = false
2026-02-02 14:12:26,132 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp7_1553_and_discretes is equal to false
2026-02-02 14:12:26,132 INFO OK: read value = false
2026-02-02 14:12:26,132 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp8_graphic_cpu is equal to false
2026-02-02 14:12:26,133 INFO OK: read value = false
2026-02-02 14:12:26,133 INFO Check that B8_MsgBitReport data_processor_test_results_DataProcessorTestResults_test_dp9_graphic_processors is equal to false
2026-02-02 14:12:26,133 INFO OK: read value = false
2026-02-02 14:12:26,133 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_array_status is equal to false
2026-02-02 14:12:26,133 INFO OK: read value = false
2026-02-02 14:12:26,133 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_cal_delta_channel_fail is equal to false
2026-02-02 14:12:26,133 INFO OK: read value = false
2026-02-02 14:12:26,133 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_cal_injection_fail is equal to false
2026-02-02 14:12:26,133 INFO OK: read value = false
2026-02-02 14:12:26,133 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_cal_noise_fail is equal to false
2026-02-02 14:12:26,133 INFO OK: read value = false
2026-02-02 14:12:26,134 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_pedestal_status is equal to false
2026-02-02 14:12:26,134 INFO OK: read value = false
2026-02-02 14:12:26,134 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_processor_status is equal to false
2026-02-02 14:12:26,134 INFO OK: read value = false
2026-02-02 14:12:26,134 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_receiver_status is equal to false
2026-02-02 14:12:26,134 INFO OK: read value = false
2026-02-02 14:12:26,134 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_rx_frontend_status is equal to false
2026-02-02 14:12:26,135 INFO OK: read value = false
2026-02-02 14:12:26,135 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_servoloop_status is equal to false
2026-02-02 14:12:26,135 INFO OK: read value = false
2026-02-02 14:12:26,135 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_test_is1_upconverter_chain_levels is equal to false
2026-02-02 14:12:26,135 INFO OK: read value = false
2026-02-02 14:12:26,135 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_test_is2_downconverter_chain_levels is equal to false
2026-02-02 14:12:26,135 INFO OK: read value = false
2026-02-02 14:12:26,135 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_test_is3_antenna_status_inconsistent is equal to false
2026-02-02 14:12:26,135 INFO OK: read value = false
2026-02-02 14:12:26,135 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_test_is4_tx_status_inconsistent is equal to false
2026-02-02 14:12:26,136 INFO OK: read value = false
2026-02-02 14:12:26,136 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_test_is5_tx_power_level is equal to false
2026-02-02 14:12:26,136 INFO OK: read value = false
2026-02-02 14:12:26,136 INFO Check that B8_MsgBitReport integrated_system_test_results_IntegratedSystemTestResults_transmitter_status is equal to false
2026-02-02 14:12:26,136 INFO OK: read value = false
2026-02-02 14:12:26,136 INFO Check that B8_MsgBitReport post_processor_test_results_PostProcessorTestResults_test_pp1_master_dsp is equal to false
2026-02-02 14:12:26,136 INFO OK: read value = false
2026-02-02 14:12:26,136 INFO Check that B8_MsgBitReport post_processor_test_results_PostProcessorTestResults_test_pp2_interface_card is equal to false
2026-02-02 14:12:26,137 INFO OK: read value = false
2026-02-02 14:12:26,137 INFO Check that B8_MsgBitReport post_processor_test_results_PostProcessorTestResults_test_pp3_cpu_cards is equal to false
2026-02-02 14:12:26,137 INFO OK: read value = false
2026-02-02 14:12:26,137 INFO Check that B8_MsgBitReport post_processor_test_results_PostProcessorTestResults_test_pp4_dma_bus is equal to false
2026-02-02 14:12:26,137 INFO OK: read value = false
2026-02-02 14:12:26,137 INFO Check that B8_MsgBitReport post_processor_test_results_PostProcessorTestResults_test_pp5_sp_interface is equal to false
2026-02-02 14:12:26,137 INFO OK: read value = false
2026-02-02 14:12:26,138 INFO Check that B8_MsgBitReport post_processor_test_results_PostProcessorTestResults_test_pp6_dp_interface is equal to false
2026-02-02 14:12:26,138 INFO OK: read value = false
2026-02-02 14:12:26,138 INFO Check that B8_MsgBitReport post_processor_test_results_PostProcessorTestResults_test_pp7_scan_converter_interface is equal to false
2026-02-02 14:12:26,138 INFO OK: read value = false
2026-02-02 14:12:26,138 INFO Check that B8_MsgBitReport post_processor_test_results_PostProcessorTestResults_test_pp8_agc_interface is equal to false
2026-02-02 14:12:26,138 INFO OK: read value = false
2026-02-02 14:12:26,138 INFO Check that B8_MsgBitReport power_supply_test_results_PowerSupplyTestResults_test_ps1_power_supply is equal to false
2026-02-02 14:12:26,139 INFO OK: read value = false
2026-02-02 14:12:26,139 INFO Check that B8_MsgBitReport power_supply_test_results_PowerSupplyTestResults_test_ps2_over_temperature is equal to false
2026-02-02 14:12:26,139 INFO OK: read value = false
2026-02-02 14:12:26,140 INFO Check that B8_MsgBitReport receiver_and_rx_frontend_test_results_ReceiverAndRxTestResults_test_fe1_lna is equal to false
2026-02-02 14:12:26,140 INFO OK: read value = false
2026-02-02 14:12:26,140 INFO Check that B8_MsgBitReport receiver_and_rx_frontend_test_results_ReceiverAndRxTestResults_test_fe2_agc_attenuators is equal to false
2026-02-02 14:12:26,140 INFO OK: read value = false
2026-02-02 14:12:26,141 INFO Check that B8_MsgBitReport receiver_and_rx_frontend_test_results_ReceiverAndRxTestResults_test_rx1_synthesizer_commands is equal to false
2026-02-02 14:12:26,141 INFO OK: read value = false
2026-02-02 14:12:26,141 INFO Check that B8_MsgBitReport receiver_and_rx_frontend_test_results_ReceiverAndRxTestResults_test_rx2_synthesizer_internal_tests is equal to false
2026-02-02 14:12:26,141 INFO OK: read value = false
2026-02-02 14:12:26,141 INFO Check that B8_MsgBitReport receiver_and_rx_frontend_test_results_ReceiverAndRxTestResults_test_rx3_uhf_oscillator_level is equal to false
2026-02-02 14:12:26,141 INFO OK: read value = false
2026-02-02 14:12:26,141 INFO Check that B8_MsgBitReport receiver_and_rx_frontend_test_results_ReceiverAndRxTestResults_test_rx4_downconverter_lo_level is equal to false
2026-02-02 14:12:26,141 INFO OK: read value = false
2026-02-02 14:12:26,141 INFO Check that B8_MsgBitReport receiver_and_rx_frontend_test_results_ReceiverAndRxTestResults_test_rx5_upconverter_lo_level is equal to false
2026-02-02 14:12:26,141 INFO OK: read value = false
2026-02-02 14:12:26,142 INFO Check that B8_MsgBitReport rx_module_test_results_RxModuleTestResults_test_rm16_calibration_sum_channel_fail is equal to false
2026-02-02 14:12:26,142 INFO OK: read value = false
2026-02-02 14:12:26,142 INFO Check that B8_MsgBitReport rx_module_test_results_RxModuleTestResults_test_rm1_master_clock_level is equal to false
2026-02-02 14:12:26,142 INFO OK: read value = false
2026-02-02 14:12:26,142 INFO Check that B8_MsgBitReport rx_module_test_results_RxModuleTestResults_test_rm2_expander_level is equal to false
2026-02-02 14:12:26,142 INFO OK: read value = false
2026-02-02 14:12:26,143 INFO Check that B8_MsgBitReport rx_module_test_results_RxModuleTestResults_test_rm3_sum_channel_down_converter is equal to false
2026-02-02 14:12:26,143 INFO OK: read value = false
2026-02-02 14:12:26,143 INFO Check that B8_MsgBitReport rx_module_test_results_RxModuleTestResults_test_rm4_dg_channel_down_converter is equal to false
2026-02-02 14:12:26,144 INFO OK: read value = false
2026-02-02 14:12:26,144 INFO Check that B8_MsgBitReport rx_module_test_results_RxModuleTestResults_test_rm5_noise_attenuators is equal to false
2026-02-02 14:12:26,144 INFO OK: read value = false
2026-02-02 14:12:26,144 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl10_agc_control is equal to false
2026-02-02 14:12:26,144 INFO OK: read value = false
2026-02-02 14:12:26,144 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl11_ad is equal to false
2026-02-02 14:12:26,144 INFO OK: read value = false
2026-02-02 14:12:26,144 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl12_das is equal to false
2026-02-02 14:12:26,145 INFO OK: read value = false
2026-02-02 14:12:26,145 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl13_serial_communications is equal to false
2026-02-02 14:12:26,145 INFO OK: read value = false
2026-02-02 14:12:26,145 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl14_taxi_interface is equal to false
2026-02-02 14:12:26,145 INFO OK: read value = false
2026-02-02 14:12:26,145 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl15_pedestal_centre_scan_location is equal to false
2026-02-02 14:12:26,145 INFO OK: read value = false
2026-02-02 14:12:26,146 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl1_low_voltage_power_supply is equal to false
2026-02-02 14:12:26,146 INFO OK: read value = false
2026-02-02 14:12:26,146 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl2_high_voltage_power_supply is equal to false
2026-02-02 14:12:26,146 INFO OK: read value = false
2026-02-02 14:12:26,146 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl3_motor_drivers is equal to false
2026-02-02 14:12:26,149 INFO OK: read value = false
2026-02-02 14:12:26,149 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl4_resolvers_power_supply is equal to false
2026-02-02 14:12:26,149 INFO OK: read value = false
2026-02-02 14:12:26,149 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl5_waveguide_switch is equal to false
2026-02-02 14:12:26,149 INFO OK: read value = false
2026-02-02 14:12:26,149 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl6_over_temperature is equal to false
2026-02-02 14:12:26,150 INFO OK: read value = false
2026-02-02 14:12:26,150 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl7_resolver_to_digital_conv is equal to false
2026-02-02 14:12:26,150 INFO OK: read value = false
2026-02-02 14:12:26,150 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl8_position_loop_error is equal to false
2026-02-02 14:12:26,150 INFO OK: read value = false
2026-02-02 14:12:26,150 INFO Check that B8_MsgBitReport servoloop_test_results_ServoloopTestResults_test_sl9_microprocessor is equal to false
2026-02-02 14:12:26,150 INFO OK: read value = false
2026-02-02 14:12:26,150 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp10_board_overall is equal to false
2026-02-02 14:12:26,150 INFO OK: read value = false
2026-02-02 14:12:26,150 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp11_attenuatori_antenna is equal to false
2026-02-02 14:12:26,151 INFO OK: read value = false
2026-02-02 14:12:26,151 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp14_external_sp_if is equal to false
2026-02-02 14:12:26,151 INFO OK: read value = false
2026-02-02 14:12:26,151 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp16_bcn is equal to false
2026-02-02 14:12:26,151 INFO OK: read value = false
2026-02-02 14:12:26,151 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp1_timer1_up is equal to false
2026-02-02 14:12:26,151 INFO OK: read value = false
2026-02-02 14:12:26,151 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp2_timer_dma_pxc_if is equal to false
2026-02-02 14:12:26,151 INFO OK: read value = false
2026-02-02 14:12:26,151 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp3_timer_internal is equal to false
2026-02-02 14:12:26,152 INFO OK: read value = false
2026-02-02 14:12:26,152 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp4_px_ctrl_comm is equal to false
2026-02-02 14:12:26,152 INFO OK: read value = false
2026-02-02 14:12:26,152 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp5_video1_without_ad is equal to false
2026-02-02 14:12:26,152 INFO OK: read value = false
2026-02-02 14:12:26,152 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp6_video1_with_ad is equal to false
2026-02-02 14:12:26,152 INFO OK: read value = false
2026-02-02 14:12:26,153 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp7_video2_ad_sync is equal to false
2026-02-02 14:12:26,153 INFO OK: read value = false
2026-02-02 14:12:26,153 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp8_video2_timer_sync is equal to false
2026-02-02 14:12:26,154 INFO OK: read value = false
2026-02-02 14:12:26,154 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp9_ad_da is equal to false
2026-02-02 14:12:26,154 INFO OK: read value = false
2026-02-02 14:12:26,154 INFO Check that B8_MsgBitReport signal_processor_test_results_SignalProcessorTestResults_test_sp9b_wideband_expander is equal to false
2026-02-02 14:12:26,155 INFO OK: read value = false
2026-02-02 14:12:26,155 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx10_hv_ps_over_temperature_warning is equal to false
2026-02-02 14:12:26,155 INFO OK: read value = false
2026-02-02 14:12:26,155 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx11_twt_helix_over_current is equal to false
2026-02-02 14:12:26,156 INFO OK: read value = false
2026-02-02 14:12:26,156 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx12_cathode_to_helix_arc is equal to false
2026-02-02 14:12:26,156 INFO OK: read value = false
2026-02-02 14:12:26,156 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx13_twt_over_temperature_hazard is equal to false
2026-02-02 14:12:26,157 INFO OK: read value = false
2026-02-02 14:12:26,157 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx14_twt_over_temperature_warning is equal to false
2026-02-02 14:12:26,157 INFO OK: read value = false
2026-02-02 14:12:26,157 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx15_cathode_under_voltage is equal to false
2026-02-02 14:12:26,158 INFO OK: read value = false
2026-02-02 14:12:26,158 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx16_cathode_over_voltage is equal to false
2026-02-02 14:12:26,158 INFO OK: read value = false
2026-02-02 14:12:26,158 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx1_microprocessors is equal to false
2026-02-02 14:12:26,158 INFO OK: read value = false
2026-02-02 14:12:26,159 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx2_tx_rf_input is equal to false
2026-02-02 14:12:26,159 INFO OK: read value = false
2026-02-02 14:12:26,159 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx3_twt_rf_input is equal to false
2026-02-02 14:12:26,159 INFO OK: read value = false
2026-02-02 14:12:26,160 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx4_twt_rf_output is equal to false
2026-02-02 14:12:26,160 INFO OK: read value = false
2026-02-02 14:12:26,160 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx5_tx_rf_output_level is equal to false
2026-02-02 14:12:26,160 INFO OK: read value = false
2026-02-02 14:12:26,161 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx6_vswr is equal to false
2026-02-02 14:12:26,161 INFO OK: read value = false
2026-02-02 14:12:26,161 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx7_three_phase_input_power is equal to false
2026-02-02 14:12:26,161 INFO OK: read value = false
2026-02-02 14:12:26,162 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx8_low_voltage_power_supplies is equal to false
2026-02-02 14:12:26,162 INFO OK: read value = false
2026-02-02 14:12:26,162 INFO Check that B8_MsgBitReport transmitter_test_results_w1_TransmitterTestResultsW1_test_tx9_hv_ps_over_temperature_hazard is equal to false
2026-02-02 14:12:26,162 INFO OK: read value = false
2026-02-02 14:12:26,163 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx17_collector_under_voltage is equal to false
2026-02-02 14:12:26,163 INFO OK: read value = false
2026-02-02 14:12:26,163 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx18_collector_over_voltage is equal to false
2026-02-02 14:12:26,164 INFO OK: read value = false
2026-02-02 14:12:26,164 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx19_rectified_voltage is equal to false
2026-02-02 14:12:26,164 INFO OK: read value = false
2026-02-02 14:12:26,164 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx20_cathode_inv_current_fail is equal to false
2026-02-02 14:12:26,165 INFO OK: read value = false
2026-02-02 14:12:26,165 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx21_collector_inv_current_fail is equal to false
2026-02-02 14:12:26,165 INFO OK: read value = false
2026-02-02 14:12:26,165 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx22_waveguide_pressurization is equal to false
2026-02-02 14:12:26,166 INFO OK: read value = false
2026-02-02 14:12:26,166 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx23_grid_window_over_duty_alt is equal to false
2026-02-02 14:12:26,166 INFO OK: read value = false
2026-02-02 14:12:26,166 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx24_floating_deck_fail is equal to false
2026-02-02 14:12:26,166 INFO OK: read value = false
2026-02-02 14:12:26,167 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx25_floating_deck_ps_fail is equal to false
2026-02-02 14:12:26,167 INFO OK: read value = false
2026-02-02 14:12:26,167 INFO Check that B8_MsgBitReport transmitter_test_results_w2_TransmitterTestResultsW2_test_tx26_grid_window_over_duty is equal to false
2026-02-02 14:12:26,167 INFO OK: read value = false
2026-02-02 14:12:26,167 INFO [Run 3] B8 Diagnostics: 173 checked, 172 pass, 1 fail
2026-02-02 14:12:26,167 WARNING B8_Degradation: 1 failures - fail
2026-02-02 14:12:26,167 INFO 0.0s remaining ...
2026-02-02 14:12:26,167 INFO [Run 3] Serial: 1 total, 0 errors, 1 fatal, 1 recycles
2026-02-02 14:12:26,168 WARNING Session: GRIFO_M_PBIT : Repetition 3 of 3
Total Step 190: Success 187 (98.42%), Failed 3 (1.58%),
2026-02-02 14:12:26,168 INFO tgt_gen()
2026-02-02 14:12:26,187 INFO TgtMsg: 1 1
2026-02-02 14:12:26,187 INFO Tgt: 1 @ 1180
2026-02-02 14:12:26,187 INFO Check that B9 b9_t_num is between 1 and 2
2026-02-02 14:12:26,188 INFO OK: read value = 1
2026-02-02 14:12:26,188 INFO Check that B9 b9_t1_rng is between 1179 and 1186
2026-02-02 14:12:26,188 INFO OK: read value = 1180
2026-02-02 14:12:26,188 INFO Repetitions terminated.
2026-02-02 14:12:26,189 INFO [Report] Custom statistics data registered for PDF generation
2026-02-02 14:12:26,190 INFO ==========================================================================================
2026-02-02 14:12:26,190 INFO FINAL TEST STATISTICS SUMMARY
2026-02-02 14:12:26,190 INFO ==========================================================================================
2026-02-02 14:12:26,190 INFO Total Runs: 3
2026-02-02 14:12:26,190 INFO Successful: 3 (100.0%)
2026-02-02 14:12:26,191 INFO Failed: 0 (0.0%)
2026-02-02 14:12:26,191 INFO B6 Checks: 36 (Pass: 30, Fail: 0, Known: 6)
2026-02-02 14:12:26,191 INFO B8 Checks: 519 (Pass: 516, Fail: 3)
2026-02-02 14:12:26,191 INFO Serial: 4 messages (0 errors, 4 fatal, 4 recycles)
2026-02-02 14:12:26,191 INFO PBIT Timing: avg=0.24s, min=0.24s, max=0.24s, σ=0.00s
2026-02-02 14:12:26,193 INFO ==========================================================================================
2026-02-02 14:12:26,193 INFO Detailed statistics will be available in the PDF report
2026-02-02 14:12:26,193 INFO Exporting statistics to CSV: C:\src\____GitProjects\SXXXXXXX_PyBusMonitor1553\__OLD\__TEST_GENOVA\GrifoAutomaticTestEnv\TestEnvironment\scripts\..\env\..\pdf_reports/GRIFO_M_PBIT\GRIFO_M_PBIT_20260202_141226_statistics.csv
2026-02-02 14:12:26,202 INFO Statistics exported successfully to: C:\src\____GitProjects\SXXXXXXX_PyBusMonitor1553\__OLD\__TEST_GENOVA\GrifoAutomaticTestEnv\TestEnvironment\scripts\..\env\..\pdf_reports/GRIFO_M_PBIT\GRIFO_M_PBIT_20260202_141226_statistics.csv
2026-02-02 14:12:26,203 INFO Statistics exported successfully to JSON: C:\src\____GitProjects\SXXXXXXX_PyBusMonitor1553\__OLD\__TEST_GENOVA\GrifoAutomaticTestEnv\TestEnvironment\scripts\..\env\..\pdf_reports/GRIFO_M_PBIT\GRIFO_M_PBIT_20260202_141226_statistics.json
2026-02-02 14:12:26,203 INFO Set MAIN_POWER to False
2026-02-02 14:12:26,203 INFO [MOCK] Power OFF
2026-02-02 14:12:26,203 INFO [MOCK] 1553 Interface stopped
2026-02-02 14:12:31,210 INFO Check that MAIN_POWER is equal to 0
2026-02-02 14:12:31,210 INFO OK: read value = False
2026-02-02 14:12:31,212 INFO [MOCK] Serial terminal disconnected
2026-02-02 14:12:31,212 INFO Stop Time: 2026-02-02 14:11:36.753001
2026-02-02 14:12:31,212 INFO Test execution time: 0:00:54.459838